Analysis of Ultra-Low Sulfur in Automotive Fuels According to ASTM D2622-10 by Sulfur Analyzer Micro-Z ULS - - Spectroscopy
 Home   Mass Spectrometry   ICP-MS   Infrared   FT-IR   UV-Vis   Raman   NMR   X-Ray   Fluorescence  
Home
Magazine
Issue Archive
Subscribe/Renew
Special Issues
Reprints
The Application Notebook
Current Issue
Archive
Submission Guidelines
Training
SpecAcademy
E-solutions
Digital Edition
Subscribe to the Digital Edition
The Wavelength
Subcribe to The Wavelength
Subscribe to the MS E-news
Resources
Market Profiles
Information for Authors
SpecTube
Webcasts
Advertiser services
Contact Us
Columns
Atomic Perspectives
Chemometrics in Spectroscopy
Focus on Quality
Laser and Optics Interface
Mass Spectrometry Forum
The Baseline
Molecular Spectroscopy Workbench

Analysis of Ultra-Low Sulfur in Automotive Fuels According to ASTM D2622-10 by Sulfur Analyzer Micro-Z ULS


Application Notebook
pp. 9

The permitted sulfur limit in fuel oils has been decreased to 10 ppm in many regions. This application note demonstrates that the performance of the Rigaku Micro-Z ULS meets the requirement of ASTM D2622-10, in the analysis of ultra-low sulfur in automotive fuels.

Recent developments in ultra-low sulfur (ULS) fuel have improved fuel efficiency and created cleaner exhaust gas. Globally, the permitted sulfur limit in fuel oils has been decreased to 10 ppm in many countries and regions.

For compliance verification, X-ray fluorescence (XRF) spectrometry is the definitive analysis tool for use at distribution terminals and refineries, as well as mobile or stationary testing laboratories. In recent years, there has been an increasing need for an instrument which does not require the use of helium gas, for instances when acquisition or delivery of helium to the analysis site is difficult.

The Micro-Z ULS (ultra-low sulfur) is a newly developed sulfur analyzer which does not require helium gas in operation. This application note demonstrates that Micro-Z ULS can meet the requirements of ASTM D2622-10.

Experimental Conditions

Instrument

The Micro-Z ULS is a benchtop wavelength dispersive X-ray fluorescence (WDXRF) spectrometer with fixed optics optimized for sulfur analysis. The atmosphere in the optics path of the Micro-Z ULS is a vacuum, so that helium gas is not required. The spectrometer is designed to minimize the restrictions in installation such as the need for cooling water, a special power supply, and installation space. The Micro-Z ULS is equipped with an air-cooled 40 W Cr-target X-ray tube and a doubly curved RX-9 analyzing crystal optimized for low concentration sulfur.

The Micro-Z ULS has a built-in operation panel, which enables simple and easy operation for daily analysis. The counting time was 300 s for peak and background measurements in this experiment.

Sample Preparation

Approximately, 4 mL of each sample was poured into a liquid cell with sample film of 2.5 Ám Mylar« film (Chemplex« 100).

Standard and Calibration


Figure 1: Calibration curve for each material of Micro-Z ULS; (a) diesel fuel, (b) isooctane (for gasoline).
"Number 2 diesel fuel" standards and isooctane-based standards provided by VHG Labs, Inc. were used for calibrations of diesel fuel and gasoline respectively.

Results

Analysis Results




Repeatability tests were carried out using a representative sample for each material. For each sample, two aliquots were prepared and analyzed using the calibration Figure 1; this process was repeated 20 times. The term "r" represents "repeatability" defined by the following formula in ASTM D2622-10 (1):

The difference between successive test results obtained by the same operator with the same apparatus under constant operation conditions on identical test material would exceed the values calculated by equation 1 only in one case in every 20.

The test results, where the maximum value for each fuel is smaller than r, prove that the performance of the Micro-Z ULS meets the requirement of ASTM D2622-10 for diesel fuel and gasoline.

Conclusion

Low concentration sulfur in petroleum-based fuel can be routinely analyzed with high accuracy and precision on the Micro-Z ULS, a benchtop WDXRF spectrometer with fixed optics, which does not require He gas, external water for cooling the X-ray tube, or special power supply. This application note demonstrates that the performance of the Micro-Z ULS meets the requirement of ASTM D2622-10, which has become strict in the recent versions of ASTM D2622.

Reference

(1) ASTM D2622-10(2010), "Standard Test Method for Sulfur in Petroleum Products by Wavelength Dispersive X-ray Fluorescence Spectrometry," ASTM International, p. 12.

Rigaku Corporation
9009 New Trails Drive, The Woodlands TX 77381
tel. (281) 362-2300, fax (281) 364-3628
Website: http://www.rigaku.com/

Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 0
Headlines from LCGC North America and Chromatography Online
Emerging Trends in Pharmaceutical Analysis
Detection of Low-Level Sulfur Compounds in Spearmint Oil
Water for GC-MS Analysis of VOCs
Differential Analysis of Olive Oils with Pegasus® GC-HRT and ChromaTOF-HRT® Reference Feature
Streamline Data Analysis of Tandem Mass Spectrometry for Inborn Errors of Metabolism Research
Source: Application Notebook,
Click here