Optimized Grazing Angle Ge-ATR of Monolayers and Thin Films on Si - - Spectroscopy
 Home   Mass Spectrometry   ICP-MS   Infrared   FT-IR   UV-Vis   Raman   NMR   X-Ray   Fluorescence  
Home
Magazine
Issue Archive
Subscribe/Renew
Special Issues
Reprints
The Application Notebook
Current Issue
Archive
Submission Guidelines
Training
SpecAcademy
E-solutions
Digital Edition
Subscribe to the Digital Edition
The Wavelength
Subcribe to The Wavelength
Subscribe to the MS E-news
Resources
Market Profiles
Information for Authors
SpecTube
Webcasts
Advertiser services
Contact Us
Columns
Atomic Perspectives
Chemometrics in Spectroscopy
Focus on Quality
Laser and Optics Interface
Mass Spectrometry Forum
The Baseline
Molecular Spectroscopy Workbench

Optimized Grazing Angle Ge-ATR of Monolayers and Thin Films on Si


Application Notebook
pp. 17

Grazing angle Ge ATR spectroscopy is extremely sensitive to monolayers and thin films on high refractive index substrates such as Si. As with any ATR method, the sensitivity increases as the critical angle is approached, as demonstrated herein.

Monolayers and other thin coatings on Si and other high refractive index substrates are of key importance in materials science, bioscience, synthetic organic, and semiconductor research. These films are often close to the FTIR detection limit so methods that improve sensitivity are vital. Traditionally, these samples were examined by transmission or grazing angle specular reflectance spectroscopy. Grazing angle Ge-ATR has recently gained wider use because of its higher sensitivity. Grazing angle Ge-ATR measurements have typically been recorded at 65, an angle sufficiently above the critical angle for the Si-Ge interface, 58.7.

This note explores the increased sensitivity gained by optimizing the incident angle and also discusses the potential spectral anomalies if the nominal critical angle of a given sample is exceeded.

Experimental


Figure 1: The VariGATR™ shown with its Force Sensor.
The grazing angle Ge-ATR measurements were carried out in a commercial FTIR spectrometer, using Harrick's VariGATR™ accessory at selected incident angles from 60 to 65 (see Figure 1). Spectra were collected at an 8 cm-1 resolution and signal averaged over 32 scans. The samples investigated were thin films on Si substrates. Each sample was pressed against the ATR crystal using the same force using the VariGATR's built-in pressure applicator and force sensor.

Results and Discussion


Figure 2: (a): ATR spectrum of sample A recorded at an incident angle of 65 (red), 63 (blue), and 60 (black) with an applied force of 722 N. (b): ATR spectrum of sample B recorded at an incident angle of 65 (blue) and 60 (black) with an applied force of 700 N.
Figures 2a and 2b show spectra recorded from two different samples at several incident angles. Both spectra have higher band intensities at the lowest angle measured. However, the spectrum of sample B shows significant dispersion effects at 60 which may complicate interpretation. These include a sloping baseline and a visible step between the high and low wavenumber edges of the bands. These effects are not apparent in the spectrum of sample A.

Conclusion

In conclusion, it is clear that selecting the incident angle to approach the critical angle does increase sensitivity. However, for ease of interpretation, the incident angle should be optimized for greatest sensitivity without significant dispersion effects.

Harrick Scientific Products, Inc.
141 Tompkins Ave., 2nd floor, Pleasantville NY 10570
tel. (914) 747-7202; fax (914) 747-7409
Website: http://www.harricksci.com/

Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 9.5
Headlines from LCGC North America and Chromatography Online
The LCGC Blog: So Just How Well Set-Up is Your UV Detector?
Agilent EU - Achiral-Chiral Heart-Cutting 2D-LC Analysis of Chiral Pharmaceutical Substances
Prosolia NA - Rapid Screening of Drugs of Abuse using Paper Spray MS
Thermo NA - Rapid Separation and Mass Spectrometry Detection of Intact Protein Samples Using Monolithic Capillary Columns
Agilent NA - Achiral-Chiral Heart-Cutting 2D-LC Analysis of Chiral Pharmaceutical Substances
Source: Application Notebook,
Click here