Moxtek's ULTRA-LITE X-ray Source


Application Notebook
pp. 11

Moxtek's ULTRA-LITE X-ray source is a very small self contained X-ray source (X-ray tube and high voltage power supply) for use in portable X-ray applications, such as the handheld X-ray fluorescence (XRF) spectrometers. This note demonstrates that this X-ray source has a stable and repeatable X-ray flux output over time, which is vital for the precision of the calibrated XRF measurements.

Moxtek's ULTRA-LITE source has a high voltage range of 5–50 kV, an emission current range of 5–200 µA, and comes with an assortment of anode materials including tungsten and silver. The maximum output power on the X-ray tube is 4 Watts. With the X-ray tube running at full power, the source draws approximately 9 Watts. At full power the ULTRA-LITE X-ray source will run for over 3 h on a typical 12 V 2.4 mA-hour batteries. It is very small in size, 25 mm × 46 mm × 148 mm, and weighs only 250 g, making it perfect for handheld and portable instruments.

Two of the key metrics of the X-ray source for XRF applications, and many other applications as well, are X-ray flux and high voltage repeatability. Repeatability means that every time the source is turned on at a particular high voltage and current setting, the high voltage and the X-ray flux stays as constant over time.

Experimental Conditions

In development of the ULTRA-LITE X-ray source we performed a number of characterization tests on multiple units. One key test is a flux repeatability test, in which the source is turned on and off numerous times. For each on/off cycle we capture an image of the focal spot, record the X-ray energy spectra coming off the tube, the energy spectra coming off a secondary XRF target, as well as input and monitor voltages and currents over several hours to days at room temperature.

Results & Conclusions


Figure 1: This shows the flux repeatability from one ULTRA-LITE source set at the maximum voltage of 50 kV and 80 µA, the full 4 Watts. The flux is stable to under 0.10% RSD (std/mean) over 3.5 days. The source was turned on for 120 s, then off for 10 s, and cycled on & off 2400 times for this test.
From these tests we have seen that the ULTRA-LITE source has a very stable and repeatable X-ray flux and high voltage. The flux was measured directly from the source with an X-ray PIN Diode, which measures the flux by counting X-ray photon events. Figure 1 shows the X-ray photon counts collected from an ULTRA-LITE source over 3.5 days of time. The PIN diode also records the energy spectra of the incoming photons, which contains the bremsstrahlung edge from the source. By tracking the bremsstrahlung edge the high voltage stability of the source was tracked over time.

The high voltage of the source does not vary more than ±100 V at 50 (0.2%) kV by measuring the bremsstrahlung edge, and the flux is stable and repeatable within 0.5% RSD or less over the time frame of days. The repeatability of this source will be an asset to any handheld or portable application where excellent flux repeatability is needed, such as XRF.

Moxtek
West 1260 North, Orem, UT 84057
tel. (800) 758-3110
Email:

Website: http://www.moxtek.com/

Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 0
Headlines from LCGC North America and Chromatography Online
Teledyne NA - Achieving ppt Levels of Environmental Volatiles with a Headspace Sampler
Thermo Fisher Scientific EU - Simplifying Complex Multi-Residue Pesticide GC-MS/MS Methodology
UCT NA - Determination of Pesticides in Strawberries Using QuEChERS Extraction, Quick QuEChERS Clean-up and GC/MS Detection
Malvern NA - Branched copolymer characterization by triple detection GPC
Malvern EU - Branched copolymer characterization by triple detection GPC
Source: Application Notebook,
Click here