Polystyrene Film as a Standard for Testing FT-IR Spectrometers - - Spectroscopy
 Home   Mass Spectrometry   ICP-MS   Infrared   FT-IR   UV-Vis   Raman   NMR   X-Ray   Fluorescence  

Polystyrene Film as a Standard for Testing FT-IR Spectrometers


Spectroscopy
Volume 28, Issue 2, pp. 38-43

Several pharmacopeias propose two tests for the resolution of Fourier transform infrared (FT-IR) spectrometers using films of polystyrene. These tests are updated versions of tests that were originally developed for dispersive instruments with the result that one of the tests is unsuitable for an FT spectrometer. In some cases, the tests are defined in absorbance, making the results dependent on sample thickness. In others, the tests are in transmission so that the results depend on the scattering properties of the sample used. Because the requirements for these properties are not defined with sufficient precision in the pharmacopeias, instruments can fail tests simply because of using an unsuitable film. Here, we illustrate these problems and suggest improvements to the tests.

Polystyrene films have long been used for checking the general performance of infrared (IR) spectrometers and specific tests for wavenumber accuracy and resolution based on such films are enshrined in various pharmacopoeias (1–4). The purpose of this article is to draw attention to some limitations of the tests for resolution proposed in several pharmacopoeias.


Figure 1: Spectra of polystyrene films supplied by instrument manufacturers. The transmission at 4000 cm-1 varies between 71% and 85% because of differences in scattering. Interference fringes are clearly visible in one spectrum and are present at lower levels in two others.
Instrument manufacturers supply polystyrene films for routine checks of their spectrometers. The films have a matte finish that reduces the amplitude of interference fringes but also increases scattering. They vary significantly in thickness, scattering properties, and even chemical composition. Spectra of a number of such films are seen in Figure 1. The differences are unimportant for wavenumber accuracy tests because the tolerances in the tests, typically ±1 cm-1, are much greater than the variations seen between different commercial polystyrene films. If required for compliance, calibrated polystyrene wavenumber standards are available from sources such as the National Institute of Standards and Technology (NIST) (5). However, the resolution tests are more problematic because they are sensitive to both the thickness and the scattering properties of the films, as well as any residual fringes. A practical consequence of this is that instruments may fail resolution checks simply because of an unsuitable choice of test film.

Experimental

Spectra were measured on a PerkinElmer Frontier FT-IR spectrometer. The resolution was 4 cm-1 except where indicated, using medium Norton-Beer apodization. Eight different films were obtained from three instrument manufacturers. The thickness of some of the films was quoted as 0.038 mm and from relative band intensities the thickness of all films was estimated to be between 0.03 and 0.038 mm. In the spectra of some films there is a band at 964 cm-1 that is associated with the presence of butadiene.


Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 4.67
Headlines from LCGC North America and Chromatography Online
Malvern NA - Branched copolymer characterization by triple detection GPC
Malvern EU - Branched copolymer characterization by triple detection GPC
Biotage EU - Fast Extraction of Metanephrine and Normetanephrine from Urine Using EVOLUTE EXPRESS CX SPE Plates
Thermo Fisher Scientific NA - New FlipBook! Clinical Research and Forensic Toxicology Applications Compendium
Pittcon 2013 Review: New Developments in Liquid Chromatography
Source: Spectroscopy,
Click here