Single reaction chamber microwave digestion enables a chemist to digest up to 15 different specialty chemical samples simultaneously
at temperatures as high as 300 °C, greatly simplifying the workflow while maintaining superior quality digestions.
 Table I: Recovery data for ERM 681
|
The class of "Specialty Chemicals" encompasses a wide variety of chemicals ranging from single chemical entities to complex
formulations. While some of these chemicals are easy to prepare for trace metals analysis, many others require extremely high
temperatures for complete digestion, which is difficult to achieve even with conventional closed vessel microwave systems.
Often chemists have to complement microwave technology with traditional tools like hot plates and Paar bombs to digest the
highly stable matrices which in turn have their own set of limitations — large acid requirements, contamination, acid handling
challenges, lengthy digestions cycles, and exposure of chemists to acid fumes. Finally, although multiple samples can be digested
in closed vessel microwaves simultaneously, samples of similar matrices need to be batched and run separately to ensure complete
control over the system. This can severely hamper the productivity of a lab testing specialty chemicals, as a wide variety
of sample matrices often require preparation.
 Table II: Recovery data for CRM8112A
|
Milestone's UltraWAVE SRC benchtop digestion system can digest up to 15 different samples simultaneously at temperatures and
pressures as high as 300 °C and 199 bar. This high temperature and pressure capability enables a complete digestion of almost
any specialty chemical that needs to be analyzed for trace metals. Samples can be directly weighed into disposable glass,
autosampler type vials with the appropriate acid mixtures; no minimum acid quantity is required. Quartz and TFM vials can
also be used depending on the user's application. This minimizes acid handling and transferring steps, which reduces errors
due to contamination and the chemist's exposure to acids. It also completely eliminates the need to clean, assemble, and disassemble
vessels used in conventional microwave systems.
Experiment
The reaction chamber was pre-pressurized to 40 bar to prevent the acids from boiling, which subsequently prevented cross-contamination
and loss of volatiles. 15-position racks were used with disposable glass vials which were covered by loose fitting caps. This
ensures pressure equilibrium on either side of the vial while preventing the condensation from the top of the chamber from
dripping into the sample. 0.2g of ERM681 (low density polyethylene) and CRM8112A (ABS Resin) were simultaneously digested
in 4 ml of 68% nitric acid. The samples were heated to 225 °C in 15 min and held at 225 °C for 15 min. The trace metals in the digested samples were analyzed using a PE Optima ICP-OES
system.
The data shows excellent DLs and recoveries. No loss of volatile elements (like Hg) or cross contamination was observed. Due
to higher sample capacity, use of disposable vials and faster cool down time, the UltraWAVE sample throughput is 2–3 times
higher as compared to closed vessel microwave digestions. Its lower consumables costs, simple operation, and superior digestion
quality make Milestone's UltraWAVE a perfect choice for specialty chemical's sample prep.