Recent advancements in photomultiplier tubes and detectors, along with improved filter contrast and transmission, have allowed
what was once impossible: detection and analysis of spontaneous movements on the nanometer level using fluorescence correlation
spectroscopy (FCS). Here, the basic techniques of FCS are discussed along with several novel applications, such as nanoparticle
dispersion studies. Insights into solving potential problems, such as photobleaching, also are provided.
Fluorescence correlation spectroscopy (FCS) is an extremely powerful, versatile technique that is useful both in vivo and
in vitro for biochemistry and biophysics. Using FCS, one can determine and obtain a great deal of quantitative information,
such as diffusion coefficients, hydrodynamic radii, kinetic chemical reaction rates, photophysical processes, and a wide variety
of internal and external dynamic characteristics of biological molecules. These measurements can be done quickly and easily
by modifying any standard fluorescence or confocal microscope. Here, we cover the important details of FCS in a few specific
examples as well as focus on the material side and see how improved filters and detectors have resurrected this once popular
technique.
How Does the Technique Work?
The technique of fluorescence correlation spectroscopy, in conjunction with confocal or two-photon excitation microscopy,
begins, as most fluorescent microscopy systems do, with focusing light on a sample. The sample's emission is measured over
time and the fluorescence intensity fluctuations are analyzed using what is called temporal autocorrelation (1). Temporal autocorrelation stems from spatial dependency and analysis of demographics or geography and is a very useful
method of measurement and analysis when working with small fluctuations in a known sample size. Essentially, it refers to
the association between time-shifted values given a voxel time series, and those particular signal values at present time
are independent of past and future signals (2). Temporal autocorrelation is important to consider because cellular and subcellular
activity is completely random and minute. Analysis of the signal during a short time domain should assist in improving the
understanding of intracellular activity. The intensity fluctuations vary depending on the sample under inspection. This is
crucial because the measured fluorescence is directly related to the intensity of the entity's overall fluctuations. In this
technique, it is very important to ensure that the number of molecules remains constant during inspection; no samples should
enter or leave the observation volume. Additionally, when the threshold of units is exceeded, fluctuations in the observation
volume become undetectable. This threshold is not a specific value but rather an observational determinant. If the inspection
region is saturated with noise and light, or if there is no noticeable flux, then the threshold more than likely has been
crossed. A consistent concentration ratio and observation volume of a variety of specimens allows researchers to conduct experiments
in a wide variety of fields, such as materials science and biology. This is not the easiest of tasks but as the saying goes,
"practice makes perfect."