Moxtek's MXDPP-50: X-ray Detection Performance Characterizations


Application Notebook
pp. 9

Moxtek now has a full energy dispersive X-ray fluorescence detector solution with Moxtek's XPINŽ detectors and our new MXDPP-50 electronics. We will present characterization methods and a few measurement details of the DPP with Moxtek's XPINŽ detectors.


Figure 1: Outlines the basic hardware for DPP and X-ray detector functional testing.
One objective for Moxtek has been to develop a DPP to work with our X-ray detector products. This has been achieved with the release of the new MXDPP-50 electronics in 2013, which comes in two versions, a boxed version (Figure 1) and a bare board OEM version.

A second objective has been to develop automated testing to fully characterize our DPP and X-ray detectors. The basic testing method is an automated test which multiple different DPP and XPIN detector settings are surveyed. The test generates a report on the performance over all the settings. This allows for rapid, consistent, and meaningful feedback, which has resulted in Moxtek more fully understanding the functionality of our DPP in development. This automated test allows for a "quick" test of our DPP and/or X-ray PIN detector over 4 to 14 standard peaking time settings, which takes about 15 to 30 min. This automated test also allows for extensive tests, takes over 16 h or more, over an array of hundreds of settings changing multiple input parameters, such as peaking time, input count rate, detector temperature, and so on.

Experimental Conditions

The basic hardware used in this automated testing is outlined in Figure 1. The DPP controls attributes such as the peaking time, the detector's temperature, the detector's high voltage, and more. The linear stage automatically positions a Fe55 isotope source to set a desired detector dead time (or detector input count rate). All the hardware is controlled in a LabVIEW program which automates all the hardware during the survey of desired settings. After each spectra is taken, over 150 independent variables are processed and recorded from both the spectra and detector, such as the Fe55 peak FWHM (for both Fe55 Kα, and Kβ), Fe55 channel position, various peak-to-background ratios, detector internal temperature, detector power draw, input count rate (ICR), output count rate (OCR), and may other variables.

Results


Table I: Outlines a small sub-set of information collected in the automated test
The automated test generates a report at the end comparing several input parameters against each other. The information generated is far greater than can be shown here (1). Table I shows a sub-set of information collected in automated test collected on a XPIN 6 mm2-BT detector at typical temperature of -35 °C and a detector dead time of 50%. To build statistically relevant results, five spectra were taken at each setting.

Many other functional dependencies are readily available as well. For example, critical variables such as the Fe55 Kα FWHM peak width can be expressed as a function of detector temperature, or the detectors count rate (1).

Conclusions

The MXDPP-50 is a digital pulse processor which now gives Moxtek a complete X-ray energy dispersive detector solution. Moxtek has been able to rigorously test the MXDPP-50, with our new suite of automated testing routines, giving a considerable amount of information quite quickly. This testing has shown the MXDPP-50 works well and works consistently from unit to unit (1).

References

(1) Link to DXC2013 poster: http://www.moxtek.com/x-ray-publications.html.




Moxtek, Inc.
452 West 1260 North, Orem UT 84057
tel. (801) 225-0930, fax (801) 221-1121
Website: http://www.moxtek.com/

Rate This Article
Your original vote has been tallied and is included in the ratings results.
View our top pages
Average rating for this page is: 0
Headlines from LCGC North America and Chromatography Online
Electronic Control of Carrier Gas Pressure, Flow, and Velocity
Application of Pyrolysis–Gas Chromatography–Mass Spectrometry for the Identification of Polymeric Materials
When Bad Things Happen to Good Food: Applications of HPLC to Detect Food Adulteration
The Column — NOW global!
Editors' Series: Gas Chromatography – Vacuum Ultraviolet Spectroscopy: A New and Worthy Alternative
Source: Application Notebook,
Click here