Inductively coupled plasma (ICP) techniques, such as ICP coupled with mass spectrometry (ICP-MS) and ICP–optical emission spectroscopy (OES), have seen a lot of growth in recent years for the direct analysis of organic samples such as petroleum and biofuels. José-Luis Todolí, a professor at the University of Alicante in Spain, has conducted several studies in this area, including the elemental determination of metals in bioethanol using ICP-OES, and the use of a torch integrated sample introduction system as well as ICP-MS to analyze petroleum products and biofuels. He recently spoke to us about this work and other projects involving ICP techniques that his group is focused on.
Inductively coupled plasma (ICP) techniques, such as ICP coupled with mass spectrometry (ICP-MS) and ICP–optical emission spectroscopy (OES), have seen a lot of growth in recent years for the direct analysis of organic samples such as petroleum and biofuels. José-Luis Todolí, a professor at the University of Alicante in Spain, has conducted several studies in this area, including the elemental determination of metals in bioethanol using ICP-OES, and the use of a torch integrated sample introduction system as well as ICP-MS to analyze petroleum products and biofuels. He recently spoke to us about this work and other projects involving ICP techniques that his group is focused on.
What do you consider to be the most challenging research application for ICP at the moment? Are you currently working on any projects that would address that challenge?
There are many challenging subjects within the field of ICP techniques. One of those is related to the direct analysis of organic samples. In this case, the petroleum and biofuels industry required approaches to overcome the problems associated with the introduction of organic solvents. In these cases, metals play an extremely important role and they should be accurately determined. Additional challenges include medical applications, nanoparticle analysis, and speciation among others. Although very important progress has been made in these fields, there is still a need for increased sensitivity and methods for overcoming interferences.
Alternatively, direct solid analysis is a subject of great interest and many different groups have been evaluating techniques such as laser ablation or laser-induced breakdown spectroscopy that require simpler and robust calibration procedures.
We have established a long-term collaboration with the IFP Energies Nouvelles in Lyon, France, to develop methodologies for improving ICP-OES and ICP-MS analytical figures of merit for the analysis of petroleum products and biofuels. This collaboration still continues, and currently we are focusing on the biofuel analysis in particular.
In this case, we are combining the advantages of using a micronebulizer at low liquid flow rates with the improvement in sensitivity encountered when a high-temperature spray chamber is employed. On the one hand, sensitivities are improved by an average factor of 2–10. This is extremely important for bioethanol samples because metals and metalloids concentrations are low in these kinds of samples. On the other hand, bioethanol may contain water at different levels depending on the particular sample. This may induce interference, thus leading to inaccurate results. By using the new device it has been possible to remove these interferences.
In an earlier paper (2), your group used the heated torch sample introduction system with ICP-MS to analyze petroleum products and biofuels. How does this method differ from the ICP-OES method discussed above? Which method is more challenging? Which method produces the best results?
Well, the system employed is basically the same. The most interesting point is that a small sample plug is injected into an air carrier stream. By working under this methodology, the sensitivity is independent of the sample nature. As a result, it is possible to analyze samples as varied as gasoline, diesel, biodiesel, and ethanol fuel by using a unique set of standards. Interestingly, it is not necessary to add an oxygen stream to avoid the formation of carbon deposits at the interface because air is introduced between consecutive injections.
Your research has spanned a wide range of application areas requiring various approaches to sample introduction. How did you get involved with developing new liquid sample introduction systems for ICP? What are the main challenges you and your group faced in developing these new systems?
The liquid sample introduction step still is one of the most critical steps in ICP techniques. The quality of the obtained results is often limited by the quality of the nebulizer, spray chamber, desolvation system, and so forth. Therefore, it is compulsory to design devices that are able to fulfill the plasma requirements (aerosol fineness, low solvent plasma load, and quick washing out) to enlarge the field of application of our technique. The main challenges faced by our research group were the plasma’s limited capability to accept a moderate solvent load without degradation in either its thermal properties or the signal stability.
You’ve also done some work with high performance liquid chromatography (HPLC) and high-temperature liquid chromatography coupled with ICP–atomic emission spectroscopy (AES) for the analysis of food products and ICP- ICP-MS for speciation (3). How did your research in this area start and what led you to develop these methods?
Coupling HPLC to ICP has been widely described. This hyphenation can supply a lot of information. Many laboratories that have an ICP instrument use it to perform elemental analyses. However, this instrument can also be used to determine the content of organic compounds. For instance, by coupling HPLC to ICP-OES it is possible to follow the signal for carbon and, hence, to access organic information. The interesting point is that by using a simultaneous ICP instrument, compounds such as sugars, carboxylic acids, alcohols, heavy metals, sulfates, and phosphates are determined together in a single chromatographic run. This kind of information is extremely useful for the food industry, for instance. Additionally, chromatographic methods, such as high-temperature LC, shorten the analysis time, which is extremely important when the employed detector is an ICP system. With regard to speciation, the use of a high-temperature mobile phase has also led to very good results. More recently, we have been developing the so-called microwave-assisted HPLC technique, which has provided interesting results in terms of analysis time, sensitivity, and resolution (4).
What are the next steps in your research?
The development and optimization of liquid sample introduction systems, including looking for more-efficient aerosol heating means, is one of our priorities. The petroleum industry is interested in this subject and the work carried out thanks to our collaboration with the IFP Energies Nouvelles will be highly useful. We are also interested in exploring the potential of laser ablation and the use of fast and accurate calibration methodologies. Finally, we are also interested in carrying out elemental speciation by means of fast chromatographic techniques using new interfaces developed in our research group.
References
Using Thermostable Raman Interaction Profiling (TRIP) For Protein Binding Screening
March 1st 2024Narangerel Altangerel, Zhenhuan Yi, and Marlan Scully of Texas A&M University recently used TRIP to analyze eight protein–ligand systems. Spectroscopy recently spoke to these three researchers about their findings and what the implications are for high-throughput drug screening.
Synthesizing Synthetic Oligonucleotides: An Interview with the CEO of Oligo Factory
February 6th 2024LCGC and Spectroscopy Editor Patrick Lavery spoke with Oligo Factory CEO Chris Boggess about the company’s recently attained compliance with Good Manufacturing Practice (GMP) International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) Expert Working Group (Q7) guidance and its distinction from Research Use Only (RUO) and International Organization for Standardization (ISO) 13485 designations.