Abstract
Silicon deposited on glass or silicon carbide is widely used in manufacturing photovoltaic cells. Both the proportion and distribution of amorphous and crystalline silicon are critical for performance and are therefore important to monitor. Raman spectroscopy is an ideal technique for this application, as the two forms generate readily distinguishable spectra that lend themselves to simple quantification methods using Beer's Law. Mapping generates chemical images with detailed information on the spatial distribution of the crystalline and amorphous forms. Since excess excitation laser power can be demonstrated to convert amorphous into crystalline silicon, care must be taken to limit the amount of power used. The Thermo Scientific DXR Raman microscope, which is equipped with a laser power regulator, is ideal for this application, particularly if the method has to be replicated from instrument to instrument at multiple manufacturing plants.
One of the more widely used photovoltaic cell technologies employs silicon deposited on either glass or silicon carbide. Although panels were made initially with either crystalline or amorphous silicon, a judicious combination of the two materials takes advantage of the benefits of each while minimizing their respective disadvantages. Optimal performance depends on reliably being able to make cells with a predetermined ratio and distribution of the two forms. Monitoring ensures the manufacture of economical, highly efficient and long-lasting solar cells.
This is an application for which Raman spectroscopy is particularly well suited. Silicon-silicon bonds are symmetrical and result in strong Raman scattering. Crystalline silicon has highly uniform bond angles and bond lengths and exists in a limited number of states. This results in sharp Raman peaks with a characteristic strong band at 521 cm–1 . Amorphous silicon is less orderly in its arrangement with a wider array of bond angles, bond energies and bond lengths in addition to dangling bonds. The distribution of possible states leads to a broad Raman band centered at 480 cm–1 that is readily distinguishable from that of crystalline silicon.
Raman spectroscopy can be used to quantify the relative amounts of amorphous and crystalline silicon in thin layer deposits. By mapping an area of deposited silicon the uniformity of the distribution of the two silicon forms can be monitored. This application note shows typical results and discusses some of the practical aspects and potential pitfalls of using Raman spectroscopy for measuring amorphous and crystalline silicon.
All spectra were collected using a Thermo Scientific DXR Raman microscope equipped with a 532 nm excitation laser, a full range grating and a motorized stage. The instrument was operated using the Thermo Scientific OMNIC 8 software suite. OMNIC™ Atlμs™ mapping software was used to collect and analyze mapping data.
Crystalline versus amorphous silicon: Figure 1 shows typical silicon spectra from samples that range from pure crystalline to those containing significant amounts of amorphous material. The strong band at 521 cm–1 of the crystalline sample very distinct from the broad band centered at 480 cm–1 that is exhibited by the amorphous silicon. These clearly distinguishable spectra make it possible to use a simple Beer's law calculation based on the ratio of the peak heights at 521 cm–1 and 480 cm–1 to estimate the relative amounts of the two forms.
Figure 1
Distribution of amorphous and crystalline silicon: Raman mapping is an excellent way to obtain information about potential variations in crystallinity over areas of deposited silicon. By mounting the sample on a motorized stage and collecting Raman spectra at specified intervals across the sample, it is possible by analyzing the spectra to determine the distribution of the silicon forms. Figure 2 shows a 30-micron line map across a sample of deposited silicon. The strong Raman band at 521 cm–1 at about half way across the map shows that there is a small region of crystalline silicon at this location. The remainder of the mapped sample shows the broad Raman band at 480 cm–1 that is characteristic of amorphous material.
Figure 2
A two-dimensional (x, y) map collected across a boundary region between crystalline and amorphous forms is shown in Figure 3. In this case OMNIC Atlµs software was used to display the spectral data as a chemical image of the distribution of the two forms by calculating the ratio of the peak heights at 521 cm–1 and 480 cm–1 . The map shows the gradient between the two forms and can be used as a quality check during manufacture.
Figure 3
Laser power: The results show that as the power of the excitation laser is increased above a threshold, amorphous silicon is converted to the crystalline form. The DXR Raman microscope used in these experiments features a laser power regulator which uses a feedback mechanism to control closely the amount of laser power delivered to the sample. This is important, as laser power output can vary quite markedly from one laser to another, even when the lasers are nominally rated to be equivalent. When using instruments without a laser power regulator, users need to exercise care, particularly when transferring methods from one instrument to another. Using the DXR Raman microscope configured with a high brightness 532 nm laser, we saw no evidence of conversion of amorphous to crystalline silicon at power levels lower than 4 mW. In developing analytical methods for this application, we recommend starting at low excitation laser power levels to ensure that no conversion of amorphous to crystalline silicon takes place.
Laser wavelength: There are three potentially important effects of excitation laser wavelength on this application: Raman efficiency, sample penetration depth and fluorescence. Since Raman scattering efficiency is proportional to 1/λ4, the signal is much stronger using short wavelength lasers and so, unless there are contraindications, it is always preferable to use a shorter wavelength.
Penetration of the laser into the silicon also diminishes with shorter excitation laser wavelengths. Thus a 532 nm laser beam penetrates approximately 0.10 μm into a silicon sample, while a 780 nm laser can penetrate up to 0.83 μm. This can be important as working with a shorter wavelength laser may be a way to minimize interference from the substrate. With silicon samples that are 100 nm or thicker there should be no interference when using a 532 nm laser. In practice it is possible to collect good data even from samples on glass that are thinner than 100 nm, since glass makes little contribution to the end result. A silicon carbide substrate will be a stronger Raman scatterer, however the Raman bands occur at a different portion of the spectrum and do not interfere with the silicon signal.
Finally, fluorescence is also excitation wavelength-dependent. Fluorescence is potentially capable of overwhelming the Raman scattering signal and should be avoided. Silicon fluoresces much more strongly at 780 nm than it does at 532 nm. Therefore fluorescence interference will be minimized when using a 532 nm laser.
As a net result, a Raman microscope for this application should be configured with a 532 nm excitation laser and should be equipped with an accurate means by which to control the laser power at the sample. A conventional neutral density filter, in the absence of a power regulator to provide feedback control, is inadequate to address the problem of laser to laser variability in power output. The laser power regulator on the DXR Raman microscope provides an ideal solution.
This work demonstrates that Raman spectroscopy is an excellent technique for monitoring deposited silicon used for photovoltaic applications. The technique readily permits quantification of amorphous and crystalline forms. Raman mapping provides detailed information about spatial distribution.
Thermo Scientific, Part of Thermo Fisher Scientific
5225 Cerona Road, Madison, WI 53711
Tel. (800) 532-4752
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.