Mercury is a toxic, persistent pollutant found in many native ores rich in copper, silver and gold. As such ores are processed mercury can be released to the environment.
Mercury is a toxic, persistent pollutant found in many native ores rich in copper, silver and gold. As such ores are processed mercury can be released to the environment. This can be either during the "stamping" process (crushing rocks to "sand") where mercury rich tailings can enter the local watershed or during smelting where volatile mercury can be released into the atmosphere. It is for these reasons and their potential impact on the environment that it is important to be able to reliably measure mercury.
Four samples of copper concentrates were acquired and analyzed using a direct analysis technique. The technique involves thermal decomposition of mercury compounds present in the sample through combustion followed by pre-concentration via gold amalgamation and subsequent detection using cold vapor atomic absorption (Figure 1).
Figure 1: Cold vapor atomic absorption illustration.
Oxygen was used as the carrier gas at a flow rate of 350 ml/min. Signal integration was set at 100 s.
The Hydra IIC provides two optical paths of differing lengths to extend its dynamic range, producing both low and high calibration curves. These calibration curves were generated using weighed deposits of aqueous standards in concentrations of 0.1, 1.0, and 10.0 ppm (w/w) mercury. All samples and standards were analyzed using nickel boats.
The low range calibration extended to 40 ng and the high range calibration to 520 ng. Both calibrations delivered linear correlation coefficients better than 0.997; however, a quadratic curve provided a slightly more precise fit for the standards.
Each of the submitted samples was introduced five times. Carryover was determined to be less than 1% by remeasuring the same sample boat with sample residue untouched. A 250 ng standard was run between sample replicates to ensure that no shift in sensitivity had occurred.
The Hydra IIC provides a 70-position autosampler and has on-the-fly loading capability for virtually unlimited capacity. Additionally, a conversion kit is available which employs chemical reduction to satisfy the monitoring of drinking water in accordance with US EPA Method 245.1 and European Standards EN1483 and EN13806.
For more information on this application or on Teledyne Leeman Lab Mercury family of products, please visit our website at www.teledyneleemanlabs.com or call 603-886-8400.
Teledyne Leeman Labs
6 Wentworth Drive, Hudson, NH 03051
tel. (603) 886-8400, fax (603) 886-9141
Website: www.teledyneleemanlabs.com
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.