Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions. If one component is volatile, trace analysis can be simplified by allowing the volatile liquids to evaporate.
Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions. If one component is volatile, trace analysis can be simplified by allowing the volatile liquids to evaporate. A multiple reflection ATR with a small sampling area is advantageous in cases where limited sample is available, as it allows additional concentration of the residue.
This note explores the use of a multiple reflection diamond ATR accessory to detect and identify the minor chemical variability between different types of nail polish remover.
Harrick Diamond ConcentratIR2™ (see Figure 1), which supplies seven reflections from the sample, was installed and aligned in a commercial FTIR spectrometer. Spectra were collected at 4 cm-1 resolution and signal averaged over 64 scans. Three nail polish removers — the original, nourishing, and strengthening formulations from the same brand — were analyzed. A single drop of each type was placed on the diamond ATR crystal. Infrared spectra were collected over a range of 4000 cm-1 to 650 cm-1 periodically until the solvent, acetone, had fully evaporated.
Figure 1: ConcentratIR2⢠Multiple Reflection ATR.
Many of the new peaks that emerge are present in all three samples. Only a few bands, in the 1600 cm-1 to 900 cm-1 region, as shown in Figure 2, are unique to any given sample. This is in good accordance with listed ingredients which indicates only a few unique substances in the analytes. Unique peaks include a H–C–H alkane bend at 1500 cm-1 , probably from tocopheryl acetate, an ingredient found only in the "nourishing" formula. Characteristic peaks are also observed at 1225 cm-1 and 985 cm-1 which are likely due to panthenol, another element particular to the nourishing formulation. Some other ingredients specific to one formula, such as citric acid (nourishing) and benzophenone-1 (strengthening) are present in quantities too small to be detected. Use of additional reflections or greater concentration of the sample might make these apparent.
Figure 2: ATR spectra of original (black), strengthening (red), and nourishing (blue) nail polish removers.
The ConcentratIR2™ can be used effectively to measure trace differences in residues of complex solutions. Since the crystal is also only 4-mm in diameter, only single drops of the sample are required, making it extremely useful in situations where only limited quantities of sample area available.
Harrick Scientific Products, Inc.
P.O. Box 277, Pleasantville, NY 10570
tel. (800) 248-3847, fax (914) 747-7209
Website: www.harricksci.com
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.