Raman spectroscopy has been limited from widespread use due to its high instrumentation cost, especially in its excitation light source. To answer this challenge, a low-cost miniature Raman spectrometer was developed. The MiniRam NIR Raman spectrometer (shown in Figure 1) utilizes B&W Tek proprietary CLEANLAZE™ laser technology and TE cooled 2048 pixel CCD array spectrometer. As a result, this low-cost instrument achieved a spectral resolution of 8 cm-1 with spectral coverage from 200 cm-1 to 2800 cm-1 .
The CLEANLAZE™ laser is a high-power broad-stripe laser diode with its output spectrum narrowed and stabilized using a novel external cavity technology. Unlike conventional external cavity lasers that require a long cavity length, the wavelength selective element in the CLEANLAZE™ laser is directly integrated onto the laser diode. As a result, the laser is both ultra compact in footprint (4 mm × 4 mm × 3 mm) and stable to environmental change. The utilization of a broad-stripe laser diode guarantees the low cost of the whole module and at the same time, it can provide much higher output power (up to 1.5 W) than its DFB and DBR counterparts. The spectral linewidth of the laser can reach 0.8 cm-1 , i.e. 0.05 nm, at 785 nm.
Figure 1. B&W Tek Mini Ram Raman spectrometer.
Utilizing a 785 nm CLEANLAZE™ laser as the Raman excitation light source, the miniature Raman spectrometer achieved a high spectral resolution and exceptional sensitivity. Other features of the Raman spectrometer include light weight (< 5 lbs), compact size (8.9" × 6.4" × 3.25"), integrated fiber optic probe, and a user friendly computer interface.
Figure 2. Raman spectrum of an amethyst sample measured with MiniRam.
The miniature Raman spectrometer has been successfully applied to a variety of applications, including gemstone identification, hazardous materials detection, pharmaceutical analysis, and epoxy cure monitoring. Some of the results are shown in Figures 2-5, respectively. Figure 2 shows the measured Raman spectrum of an amethyst sample. The MiniRam provides a quick, non-destructive gemstone identification method from their characteristic Raman spectra. Figure 3 shows the measured Raman spectrum of a Benzenethiol sample (a hazardous material) that was captured on a Mesophotonics Klarite SERS substrate (shown in the upper right corner of Figure 1).
Figure 3. SERS spectrum of a Benzenethiol sample measured with MiniRam.
The surface enhancement provided by the substrate and the high sensitivity of the MiniRam allow us to detect Benzenethiol at low concentration levels. Figure 4 shows the measured Raman spectrum of a Tylenol™ sample. The high spectral resolution of the Raman spectrum provides a convenient non-contact way to analyze the content and active ingredient distribution of the medicine. Figure 5 shows the Raman spectra of an epoxy sample measured in real time with MiniRam during the cure process. The variation of the relative intensity of the 1260 cm-1 epoxy band was used to accurately monitor the curing status of the epoxy sample.
Figure 4. Raman spectrum of a Tylenol sample measured with MiniRam.
A miniature Raman spectrometer utilizing spectrum narrowed diode laser was developed and applied to a variety of applications. The measured Raman spectra show good spectral resolution and sensitivity for detection, identification, analysis and monitoring of a target material.
Figure 5. Raman spectra of an epoxy sample measured in real time with MiniRam during the cure process.
B&W Tek, Inc.
19 Shea Way, Suite 301, Newark, DE 19713
Tel. (302) 368-7824, Fax (302) 368-7830
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
Accurate and robust analysis of challenging environmental samples for regulatory compliance
April 25th 2024This application note discusses how the adverse effects of sample matrix can be circumvented with an analytical workflow that includes robust ICP-MS instrumentation offering comprehensive yet simplified analysis of samples containing high dissolved solids such as soil digests, wastewater and solid wastes. The data presented in this application note was generated over 12 hours with over 160 samples covering soil extracts and a variety of water samples.
Analysis of drinking water in compliance with EPA Method 200.8, Revision 5.4
April 25th 2024This application note describes the analytical workflow developed for the analysis of water samples over an extended period using the latest innovations in ICP-MS instrumentation. The innovations specifically highlighted in this app note are the next generation Argon Gas Dilution accessory, which simplifies the analysis of high matrix samples allowing direct analysis without prior manual diution; the HAWK consumables and maintenance assistant within the Qtegra ISDS software; the new Easy-click compact (ECC) peristaltic pump; and the new Thermo Scientific autosampler, the iSC-65 autosampler.