A Corrosive Liquid Investigated by Diamond ATR Infrared Spectroscopy

Experimental

Infrared spectra were collected on an FT-IR spectrometer equipped with the Harrick DiaMaxATR™ single-reflection high-throughput diamond ATR accessory (see Figure 1). The system was purged to remove water vapor and CO2. Spectra were collected at 4 cm-1 resolution and the signal was averaged over 16 scans. The spectra were referenced to the clean ATR crystal.

The sample used here was Easy-Off™ Heavy Duty Oven Cleaner (Reckitt Benckiser). Two drops of the oven cleaner were placed on the center of the ATR crystal and spectra were collected at 20 scans per s over a period of 60 min.

Results and Discussion

Figure 2 shows the series of FT-IR spectra acquired over the period of 60 min, showing water evaporating from the oven cleaner and concentrating the remaining ingredients. The MSDS for the oven cleaner indicates that it is composed of 2-(2-Butoxyethoxy)ethanol, 2-amino-ethanol, sodium hydroxide, and liquefied petroleum gases. As it dries, the OH stretching and bending vibrations around 3300 and 1645 cm-1 respectively decrease, showing the evaporation of water present with sodium hydroxide. The bands at 2958, 2930, 2862, and 2800 cm-1, due to CH stretches, increase as the oven cleaner becomes more concentrated, as do the amine bands at 1658 and 1590 cm-1. Presumably the volatile petroleum components are also evaporating. But the decrease in the volatile organics is more difficult to discern, due to the increased concentration of the nonvolatile compounds. Note that the rate at which the oven cleaner concentrates could be measured by looking at peak areas, as demonstrated by the inset graph for the C=C transition at 1000 cm-1



Conclusion

Because diamond is a more chemically durable material, it can be effectively used as an ATR crystal for some corrosive samples. This has been demonstrated here using the Harrick DiaMaxATR to measure the ATR spectra of oven cleaner, as it becomes increasingly concentrated due to evaporation of the solvent therein.

Harrick Scientific Products, Inc.
141 Tompkins Ave., 2nd Floor, Pleasantville, NY 10570
tel. (800) 248-3847
Website: www.harricksci.com


Related to this article

Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.
Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.