A Diamond ATR-FTIR Study of Nitriles

This applications note compares the sensitivity of a single-reflection ATR to multiple-reflection ATR for the nitrile functional group infrared transition.

Diamond ATR has become one of the most commonly used FT-IR spectroscopy methods. However, the strong diamond lattice bands in the 2300–1900 cm-1 region make it difficult to measure the functional groups from nitriles, isocyanates, isothiocyanates, diimides, azides, and ketenes that would normally appear in that region. This applications note compares the sensitivity of a single-reflection ATR to multiple-reflection ATR for the nitrile functional group infrared transition.

Experimental

Infrared spectra were collected on an FT-IR spectrometer equipped with the Harrick DiaMaxATR™ single-reflection high-throughput diamond ATR accessory or ConcentratIR2 multiple-reflection diamond ATR (see Figure 1). The system was purged to remove water vapor and CO2. Spectra were collected at 8 cm-1 resolution, and signal averaged over 32 scans. The spectra were referenced to the clean ATR crystal.


Figure 1: Two diamond ATRs: Single reflection DiaMaxATR (left) and ConcentratIR2 (right).

The sample used here was a High Five Nitrile Disposable Glove (lightly powdered, textured, P/N N842). For the single-reflection ATR measurement, a portion of the glove was pressed against the ATR crystal using the maximum force supplied by the build-in pressure applicator. For the multiple-reflection ATR measurements, the sample was gently pressed against the ATR crystal with compressible foam to fill the trough, and to apply force as uniformly as possible.

Results and Discussion

Figure 2 shows the two spectra. The upper spectrum measured with the multiple-reflection ATR shows only noise in the 2300–1900 cm-1 region, due to the high absorbance of diamond in that region. The lower spectrum, however, clearly shows the C≡N stretch at 2238 cm-1. Another weaker band is also apparent in the single-reflection ATR spectrum at 2162 cm-1, possibly indicating the presence of an alkyne group.


Figure 2: Diamond ATR spectra of a nitrile glove measured using a multiple reflection ATR (blue) and a single reflection ATR (red).

Conclusion

The Harrick DiaMaxATR high throughput, single reflection diamond ATR is effective at measuring functional groups that are frequently obscured by the strong diamond lattice bands in the 2300–1900 cm-1 region. This makes it more suitable for analysis of the function groups in that region, including nitriles, isocyanates, isothiocyanates, diimides, azides, and ketenes.

Harrick Scientific Products, Inc.
141 Tompkins Ave., Box 277, Pleasantville, NY 10570
tel. (914) 747-7202, fax (914) 747-7409
Website: www.harricksci.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.
Highlights from What’s Nu October 2026
In this Q&A, we recap the October edition of “What’s Nu,” which covered topics such as how novel spectroscopic and analytical methods are transforming environmental monitoring, medical treatment, and accessibility in scientific research.