A team of researchers has successfully developed 3D printed multi-hole collimators for X-ray imaging systems, offering a cost-effective alternative to traditional pinhole apertures. This development has the potential to enhance the sensitivity and position resolution of X-ray fluorescence (XRF) spectrometers, paving the way for more accurate and detailed analyses.
The study, published in Spectrochimica Acta Part B: Atomic Spectroscopy, explored the performance of multi-hole collimators with varying geometries and material compositions in a full-field XRF spectrometer based on the 2D-THCOBRA gas detector (a special micropattern gaseous detector [MPGD] design). The researchers evaluated the relative sensitivity and position resolution of different setups, comparing them to pinhole-based configurations (1).
Their findings revealed that the use of smaller pinholes, while limiting system sensitivity, provides superior position resolution. Conversely, collimators with large holes and small septa yielded higher sensitivity but compromised position resolution. The best position resolution, 1.10 ± 0.02 mm full width at half maximum (FWHM), was achieved with a 0.3 mm diameter pinhole aperture, which also exhibited the lowest relative sensitivity. The lead (Pb) collimator with 2 mm hexagonal holes presented the worst position resolution value, 3.25 ± 0.09 mm FWHM (1).
In terms of sensitivity, the low attenuation coefficient and small depth of the copper (Cu) collimator with 1 mm circular holes resulted in the highest sensitivity–50 times greater than the value achieved with the smaller pinhole aperture.
These findings underscore the critical trade-off between sensitivity and position resolution in XRF imaging, emphasizing the importance of tailoring the optical component to the specific application. While pinhole apertures allow magnification for enhanced detail, they may compromise sensitivity. Multi-pinhole collimators, as used in SPECT imaging of small animals, offer a balance between sensitivity and position resolution.
The researchers highlight the potential of 3D printed collimators as a cost-effective and versatile solution for energy dispersive X-ray fluorescence (EDXRF) system applications. 3D printing enables the exploration of diverse geometries and designs, paving the way for further advancements in XRF imaging technology.
This research holds significant implications for various fields, including environmental monitoring, material analysis, and biomedical imaging. The development of 3D printed multi-hole collimators promises to enable more accurate, sensitive, and versatile analyses.
This article was written with the help of artificial intelligence and has been edited to ensure accuracy and clarity. You can read more about our policy for using AI here.
Reference
Carvalho, P. M. S.; Leite, F. D.; Tavares, G.; Correia, P. M. M.; Oliveira, J. M. M.; Veloso, J. F. C. A.; Silva, A. L. M. 3D Printed Imaging Apertures for MicroPattern Gas Detector Based Full Field of View X-Ray Fluorescence Spectrometers. Spectrochim. Acta Part B: Atomic Spectrosc. 2023, 210, 106806. DOI: 10.1016/j.sab.2023.106806.
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.
Raman Reveals Rare Ribbeck Meteorite Clues About Ancient Solar System
May 28th 2025A team of scientists in Poland has unveiled the first detailed structural and magnetic analysis of the Ribbeck meteorite, a recently recovered space rock classified as an aubrite. Using Raman spectroscopy, X-ray diffraction, and advanced magnetic testing, researchers revealed the meteorite's unique mineralogy and its connection to deep space conditions.
Multi-Analytical Study Reveals Complex History Behind Ancient Snake Motif in Argentine Rock Art
May 22nd 2025A recent study published in the Journal of Archaeological Science: Reports reveals that a multi-headed snake motif at Argentina's La Candelaria rock shelter was created through multiple painting events over time.
Using Spectroscopy to Understand Hawaii’s Hidden Geology
May 8th 2025Researchers from the University of Nevada, Reno, have conducted the most comprehensive subsurface analysis of Hawaiian shield basalts to date, using advanced spectroscopic and geochemical techniques to reveal short-lived hydrothermal alteration processes and establish a new foundation for future volcanic and geothermal studies.