Tuning the Photoluminescence of Graphene Oxide

The emission wavelength of graphene oxide is tuned through photothermal reduction and the spectral shift monitored using photoluminescence spectroscopy to achieve the desired emission.

The emission wavelength of graphene oxide is tuned through photothermal reduction and the spectral shift monitored using photoluminescence spectroscopy to achieve the desired emission.

Graphene has received tremendous attention due to its remarkable electrical, thermal, and mechanical properties. However, the valence and conduction bands of graphene touch, and pure graphene, therefore, has zero bandgap, and does not exhibit photoluminescence (PL). When oxidized to form graphene oxide (GO), the π electron network of graphene is disrupted, which induces a bandgap, and the material becomes photoluminescent. This opens the door to optoelectronic applications, and is investigated using PL spectroscopy in this application note.

Experimental

An aqueous dispersion of GO (PlasmaChem GmbH) was diluted to 0.02 mg cm-3 and the PL measured using the FLS1000 Photoluminescence Spectrometer equipped with double excitation and emission monochromators and PMT-900 and PMT 1700 photomultiplier tube detectors.

Results

The high scatter and low quantum yield of GO makes it a challenging sample for most PL spectrometers to measure because the monochromators are unable to completely remove the scattered excitation light. The FLS1000 can be equipped with double monochromators, which increases the stray light rejection of the monochromator from 1:1×105 to 1:1×1010 and ensures the spectra of even the most weakly emitting and highly scattering samples can be measured accurately. The PL spectrum of GO was measured using double monochromators and extends from 350 nm to 1250 nm with a peak at 710 nm (Figure 1 inset).


Figure 1: Change in the PL spectra of GO with increasing photothermal reduction. λex = 300 nm, Δλex = 5 nm, Δλem = 5 nm. Inset: complete spectrum.

To tune the emission wavelength of GO, it was photothermally reduced by exposure to a Xenon lamp and the PL spectra measured after different exposure times (Figure 1). As the exposure time is increased, the PL blue shifts with a maximum at 450 nm after 60 min. The blue shift arises from the relative intensity of the 450-nm peak, increasing at the expense of the 710-nm, which reveals that the peaks correspond to two chemically distinct species, with the 710-nm being converted to the 450-nm species during reduction.

Conclusion

The PL of GO was investigated using the FLS1000 Photoluminescence Spectrometer. Pristine GO was found to exhibit broad PL with a maximum at 710 nm. By photothermally reducing the GO and monitoring the change in emission with the FLS1000, the PL could be tuned from 710 nm to 450 nm.

References

(1) A. Sehrawat, P. Sehrawat, S. S. Islam, P. Mishra, and S. Ahmad, Sci. Rep.8, 3537 (2018).

(2) C. Chien, S. Li, W. Lai, Y. Yeh, H. Chen, I. Chen, L. Chen, K. Chen, T. Nemoto, S. Isoda, M. Chen, T. Fujita, G. Eda, H. Yamaguchi, M. Chhowalla, and C. Chen, Angew. Chem. Int. Ed. 51, 6662 (2012).

Edinburgh Instruments
2 Bain Square, Kirkton Campus
Livingston, EH54 7DQ, United Kingdom
tel. +44 (0) 1506 425 300
Website: www.edinst.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.
Highlights from What’s Nu October 2026
In this Q&A, we recap the October edition of “What’s Nu,” which covered topics such as how novel spectroscopic and analytical methods are transforming environmental monitoring, medical treatment, and accessibility in scientific research.