Fast Quality Evaluation of Bulk Polymers and Products Using FT-NIR Spectroscopy

FT-NIR spectroscopy for the quantitative determination of key physical parameters in bulk polymers and products, including density and melt index (MI). These measurements can often be made with little or no sample preparation.

FT-NIR spectroscopy for the quantitative determination of key physical parameters in bulk polymers and products, including density and melt index (MI). These measurements can often be made with little or no sample preparation.

Fourier transform (mid) infrared spectroscopy is an established technique for the fast and nondestructive analysis of plastic films, where it can typically provide critical compositional information, such as additive content and co-polymer concentration. However, in the manufacturing processes for bulk polymers of polyethylene (PE), physical property determination (for example, density and melt index) can be important quality control parameters. Due to the simplicity of sampling when using near infrared spectroscopy, FT-NIR has proved to be a very adaptable and useful method in these circumstances.

Experimental Conditions

For the PE samples, 65 different bulk polymer samples were provided, each sample in both pellet and powder form, so a total of 130 sample measurements were made. Both pellet and polymer samples were presented in glass scintillation vials approximately 2 cm in diameter and 6 cm in height, and were loaded into the FT-NIR spectrometer rotating powder sampler device without further sample preparation and measured in diffuse reflectance.

Figure 1: PE pellet samples - variation in FT-NIR spectra as MI and density change.

Spectral acquisitions were performed using an ABB MB3600 FT-NIR spectrometer, fitted with a TE-cooled InGaAs detector at 16 cm-1 resolution and with 128 co-added scans. The background reference measurement was made using a PTFE standard.

Calibration models were developed using PLS/IQ (Grams/AI version 7.0 software). Calibration model uncertainty is estimated by the standard error of cross validation (SECV). The static repeatability of the calibration is given by the standard deviation of ten consecutive measurements of the same sample without moving it from the sampling accessory.

Results and Discussion

The variations of properties in this study are presented in Figure 1.

Figure 2 displays the vibration regions of -CH3 or -CH2 function groups and spectra variation as to the change of properties in PE pellets samples.

Figure 2: Experimental design - variation of MI and density in PE sample set.

The calibration results are summarized in Table I for PE samples. It has been demonstrated that ABB MB3600 with a Powder SamplIR accessory could successfully determine MI and density in pellet and powder PE samples with: 1) SECV< lab reproducibility, 2) FT-NIR repeatability < lab repeatability.

Table I: Calibration results

Conclusion

FT-NIR technology could provide the solutions for the realization of at-line or on-line determination of these properties and provide quick feedback to the plant operator for the quality control.

ABB Analytical

585 Blvd. Charest E., Ste 300, Quebec, QC Canada G1K9H4

tel. 418 877 2944 Email: ftir@ca.abb.com; Website: www.abb.com/analytical


Related to this article

A robotic arm in an automated agricultural setting, carefully processing a conveyer belt with tomatoes and leafy greens for food production. | Image Credit: © Donnangelo - stock.adobe.com
Croatian researchers outline how optical emission spectroscopy (OES) can track reactive species and plasma parameters in real time, giving processors a path to move cold plasma treatments from the laboratory to production lines.
Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.