Real-Time, Broad-Spectrum Odor Analysis Using SIFT-MS

The sensitive, selective, and real-time analysis characteristic of the SIFT-MS technique provides simple, robust, and continuous analysis of extremely diverse odor compounds at trace levels in air. This application note illustrates instantaneous, broad spectrum odor analysis with monitoring data from a chicken meat production facility.

The sensitive, selective, and real-time analysis characteristic of the SIFT-MS technique provides simple, robust, and continuous analysis of extremely diverse odor compounds at trace levels in air. This application note illustrates instantaneous, broad-spectrum odor analysis with monitoring data from a chicken meat production facility.

Odor compounds are chemically very diverse, making comprehensive analysis very challenging using conventional analytical technologies. Furthermore, odors tend to be dynamic, requiring fast response, whereas odor panels or lab-based analyses are expensive and based on time-averaged samples.

Alternatively, selected ion flow tube mass spectrometry (SIFT-MS) simultaneously detects and quantifies both organic and inorganic odorous compounds in real-time in air to pptv levels with no sample preparation (1,2). SIFT-MS is a direct mass spectrometry technique that utilizes multiple, rapidly switchable chemical ionization agents to deliver unparalleled real-time selectivity, absolute quantitation, and wide linear and dynamic ranges.

This application note illustrates how readily a SIFT-MS instrument can be applied on-site for comprehensive real-time odor analysis.

Experimental Conditions

A van-mounted Syft Technologies’ Voice200 SIFT-MS instrument was deployed at several chicken production facilities in South East Queensland, Australia, during September and October 2013.

Air from the poultry shed was sampled continuously by the SIFT-MS instrument using a flow-past configuration. A sampling pump drew air at a flow rate of a few liters per min through Teflon tubing and the SIFT-MS subsampled it at a flow rate of 25 sccm.

A diverse range of odorous compounds was targeted based on a study using GC–MS (3), augmented with other odorous compounds that are difficult to detect using traditional chromatographic methods (for example, ammonia and hydrogen sulfide).

Results

Figure 1 shows a 13-h snapshot of real-time data obtained while chickens were harvested and the shed cleared of used litter. A diverse range of odor compounds were detected and quantified, including ammonia, amines, hydrogen sulfide, organosulfur compounds, ketones, aldehydes, and organic acids. The odor profile changes dramatically over the analysis period, illustrating the value of continuous, broad-spectrum monitoring using SIFT-MS compared to time-averaged methods.

Figure 1: Comprehensive, continuous odor monitoring using SIFT-MS during harvesting of meat chickens and the subsequent clean out of the production shed.

 

Conclusions

The real-time, broad-spectrum analysis provided by SIFT-MS makes it ideally suited to continuous monitoring of odorous organic and inorganic compounds even at the very low concentrations required for investigation of odor complaints. The Syft Voice200ultra SIFT-MS solution provides a robust, easily deployed and operated package for on-site monitoring, but is equally at home in contract and research laboratories.

References

  • B.J. Prince, D.B. Milligan, and M.J. McEwan, Rapid Commun. Mass Spectrom. 24, 1763–1769 (2010).

  • V.S. Langford, I. Graves, and M.J. McEwan, Rapid Commun. Mass Spectrom. 28, 10–18 (2014).

  • K.R. Murphy, G. Parcsi, and R.M. Stuetz, Chemosphere95, 423–432 (2014).

 

Syft Technologies

3 Craft Place, Christchurch 8024, New Zealand

tel. +64-3-338 6701, fax +64-3-338 6704

Website: www.syft.com

 


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.