TMD Layer Characterization with Raman and RISE Microscopy

Author(s)WITec GmbH

This study describes how Raman imaging and correlative Raman scanning electron (RISE) microscopy can characterize and visualize layer numbers in 2D MoS2 and WSe2 samples.

Transition metal dichalcogenides (TMDs) exhibit unique electronic and optical properties, and these differ depending on the number of layers of material. Raman imaging can quantify layers by these differences. Raman microscopy can also be combined with other techniques, such as scanning electron microscopy, for a more comprehensive view.

Experimental

An investigation of a tungsten diselenide flake quickly produced high-resolution images with a WITec alpha300 apyron fully automated Raman microscope.

Molybdenum disulfide monolayer crystals were examined with a WITec RISE system featuring a common vacuum chamber for both Raman and SEM measurements.

Results

Ultrafast Raman Analysis of Tungsten Diselenide (WSe2)

Figure 1B shows a measurement of a WSe2 flake that was acquired in approximately 2 min. It is a clear and detailed 75 × 75 µm² Raman image consisting of 10,000 spectra. The same data after smoothing are also shown. A measurement made in about 17 min of 100,000 spectra produced an even sharper image.

RISE Microscopy Investigation of Molybdenum Disulfide (MoS2)

RISE microscopy reveals structure as well as crystalline and exciton dynamics of few-layered transition metal dichalcogenides (TMDs). Twin crystals of MoS2 on SiO2/Si appear in the SEM image as star-shaped forms (Figure 2A). The Raman spectra of MoS2 monolayers show the characteristic E’/E2g and A’1/A1g Raman band modes of MoS2 at 385 and 403 relative wavenumbers (cm-1) respectively (Figure 2C). With an increasing number of layers, the two Raman bands drift apart due to in-plane and out-of-plane (inter-layer) vibrations.

The grain boundaries visible in the SEM image correlate with the areas where Raman signals indicate crumpled or overlapping edges, highlighting the effectiveness of RISE microscopy (Figure 2B) for characterizing MoS2 crystals.

Conclusion

Raman microscopy and the combination of Raman imaging with scanning electron microscopy offer researchers fast and versatile tools for determining the number of layers in transition metal dichalcogenide samples. High-resolution images of tungsten diselenide and molybdenum disulfide crystals can show edges, boundaries, folds, wrinkles, and other structures in clear, precise detail. This capability will be of great use as 2D materials and TMD research continues.

WITec GmbH
Lise-Meitner-Str. 6, 89081 Ulm, Germany
tel. +49 731 140 70 0, fax +49 731 140 70 200
Website: www.witec.de


Related to this article

Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.
Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.