Moxtek's MAGPRO miniature X-ray source is capable of up to 60 kV 200 µA and down to 4 kV and 1000 µA, and limited to an output power of 12 watts. The 12 W source is meant for use in small bench top systems where small size is valued.
The 12 W miniature source has a higher voltage and a much higher emission current than our present miniature X-ray sources, which is an advantage for a number of applications. The initial release of the 12 W source, our MXDPP-50 electronics comes in two versions: The tungsten anode version, which can be used in either XRF or in X-ray imaging, and the Rh anode version, primarily meant for light element XRF detection (Figure 1). The entire X-ray source (tube plus high voltage power supply) weighs about 700 g (25.0 oz) and can be controlled by your choice of USB, digital, or analog control. It also has an interlock connection for X-ray safety.
We will show the advantage of the rhodium anode source running at 4 kV 1000 µA for XRF of cement powders.
Concentrations of sulfur (sulfate), silicon (silicon dioxide), and aluminum (aluminum oxide) are important for both identifying and controlling different blends of cement powder. XRF can be used to measure cement powder elemental concentrations. XRF for the desired "light" elements, with emission lines under about 4 keV is challenging because the characteristic X-rays are absorbed by air, thereby attenuating the signal from the sample. For this experiment, we used a typical XRF setup where the sample is placed at 45° to the tube, and 135° to the detector, and no filters on the 12 W X-ray source. For light element XRF, a rhodium anode is often used because it has L emission lines between 2.7 and 3.0 keV, which are used to excite elements with emission lines below 2.5 keV. For this test the 12 W X-ray source was set to the minimum of 4 kV, which is helpful in eliminating the calcium XRF signal from the cement. The calcium signal is unwanted in this application because it will overwhelm the detection of the other elements of interest. A high emission current X-ray source, such as 1000 µA from the 12 W source, is very helpful in increasing the signal.
Figure 2 shows an XRF spectrum from a Portland cement sample with the 12 W source running at 1000 µA and 200 µA on a rhodium anode tube. The signal level is increased by a bit over fivefold for these light elements, which can be a huge advantage in any application which detection of light elements is needed.
The new 12 W source is meant for use in small benchtop systems where small size is valued. It has as increased range up to 60 kV at high voltage and increased current to 1000 µA at low voltages; both of which are an advantage in XRF and imaging applications.
Moxtek
452 West 1260 North, Orem, UT 84057
tel. (800) 758-3110, email: info@moxtek.com
Website: www.moxtek.com
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.
Specificity and the Net Analyte Signal in Full-Spectrum Analysis
July 21st 2025This tutorial addresses the critical issue of analyte specificity in multivariate spectroscopy using the concept of Net Analyte Signal (NAS). NAS allows chemometricians to isolate the portion of the signal that is unique to the analyte of interest, thereby enhancing model interpretability and robustness in the presence of interfering species. While this tutorial introduces the foundational concepts for beginners, it also includes selected advanced topics to bridge toward expert-level applications and future research. The tutorial covers the mathematical foundation of NAS, its application in regression models like partial least squares (PLS), and emerging methods to optimize specificity and variable selection. Applications in pharmaceuticals, clinical diagnostics, and industrial process control are also discussed.
New Study Expands Nickel Autoionization Spectra to Advance Laser Isotope Separation Technologies
July 17th 2025Researchers at China’s National Key Laboratory have identified 170 nickel autoionization states using resonance ionization mass spectrometry, significantly advancing the spectral database critical for laser isotope separation and atomic spectroscopy.
AI-Powered Fusion Model Improves Detection of Microplastics in the Atmosphere
July 17th 2025Researchers from Nanjing University of Information Science & Technology have introduced a breakthrough AI-enhanced multimodal strategy for real-time detection of polyamide microplastics contaminated with heavy metals.
How Analytical Chemists Are Navigating DOGE-Driven Funding Cuts
July 14th 2025DOGE-related federal funding cuts have sharply reduced salaries, lab budgets, and graduate support in academia. Researchers view the politically driven shifts in priorities as part of recurring systemic issues in U.S. science funding during administrative transitions. The impact on Federal laboratories has varied, with some seeing immediate effects and others experiencing more gradual effects. In general, there is rising uncertainty over future appropriations. Sustainable recovery may require structural reforms, leaner administration, and stronger industry-academia collaboration. New commentary underscores similar challenges, noting scaled-back graduate admissions, spending freezes, and a pervasive sense of overwhelming stress among faculty, students, and staff. This article addresses these issues for the analytical chemistry community.