FT-IR Sampling of Residues on Curved Metal Surfaces Outside of the Sampling Compartment

In the manufacturing process, it frequently becomes important to determine if metal parts are clean or are sufficiently lubricated with oil. Infrared analysis of small, flat pieces can be readily carried out by in-compartment grazing angle specular reflectance.

Susan Berets and Jeffrey Christenson, Harrick Scientific Products, Inc.

In the manufacturing process, it frequently becomes important to determine if metal parts are clean or are sufficiently lubricated with oil. Infrared analysis of small, flat pieces can be readily carried out by in-compartment grazing angle specular reflectance. However, parts are frequently large and irregularly shaped or curved, which makes in-compartment analysis trickier. This note explores the use of a specular reflectance fiber optic probe to measure oil on the curved surface of a large metal cylinder outside of the spectrometer sampling compartment.

Experimental

The measurements were carried out using Harrick Omni-Spec fiber optic specular reflectance probe with two PIR fibers (Figure 1), interfaced to a commercial FT-IR spectrometer through the Harrick FiberMate2 coupler. Spectra were collected at an 8 cm-1 resolution and signal averaged over 32 scans with an MCT-A detector. The sample investigated was a stainless steel cylinder, 6 in. in diameter and 4.5 in. high, with a machined finish. The sample was cleaned for the background measurement and a thin coating of Mazola oil was applied to the same area prior to the sample measurement. For both the sample and background measurements, the Omni-Spec was held so its sampling surface was tangential to the surface of the sample.

Figure 1: Omni-Spec specular reflection fiber optic probe.

Results and Discussion

Figure 2 shows the oil coated metal. As expected, there are C-H stretching bands prominent in the 2800–2900 cm-1 region, in addition to bands at 1724 cm-1 due to a C=O stretch, 1462 cm-1 due to C-H bends, and 1160 cm-1 from a C-O stretch. Note that there is increased noise about 3000 cm-1 . This is due to the optical cut-off of the PIR fiber. The baseline of the spectrum is also strongly sloped. This is due the machine finish on the sample which creates shiny and rough areas on the surface. The Omni-Spec collects the specular reflectance in addition to some fraction of the diffused specular reflectance. This combination results in a sloping baseline.

Figure 2: Specular reflectance spectrum of oil on surface of machined steel cylinder.

Conclusion

Specular reflection fiber optic probes such as the Omni-Spec can be used effectively to examine coatings and residues on large, curved metal surfaces, provided that care is taken in positioning the probe relative to the sampling surface. The Omni-Spec fiber optic probe operates outside of the spectrometer sample compartment, making it a useful tool for examining large samples.

Harrick Scientific Products, Inc.

141 Tompkins Ave., Box 277, Pleasantville, NY 10570

tel. (914) 747-7202, fax (914) 747-7409

Website: www.harricksci.com


Related to this article

Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.
Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring-bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.