Impact of Containers on Raw Material Verification Using a Handheld Raman Spectrometer

Raman spectroscopy is a well-accepted, powerful tool for material identification, and the ability of a Raman spectrometer to acquire spectra through plastic and glass packaging materials protects sample integrity and greatly reduces analysis time.

Raman spectroscopy is a well-accepted, powerful tool for material identification, and the ability of a Raman spectrometer to acquire spectra through plastic and glass packaging materials protects sample integrity and greatly reduces analysis time.

Figure 1

Experiment

A TruScan handheld Raman spectrometer from Ahura Scientific was used in this work. To verify sample identity, TruScan calculates a p-value, which is the probability that differences between the sample and reference spectra are due to the uncertainty of the measurement rather than to differences in molecular structure. If the p-value ≥ 0.05 a PASS result is displayed. For p-values < 0.05 a FAIL result is returned. A single reference spectrum for each material was acquired "neat," with no container wall between the sample and laser aperture of the spectrometer. Three measurements were made for each material and packaging combination listed in Figure 1.

Figure 2

Results

In Figure 1, green indicates p ≥ 0.1, well above the 0.05 PASS threshold. While results in the range of 0.05 < p-value < 0.1 (green with hash marks) are above the PASS threshold, there may be a higher probability of false-negative results. For those bottle containers, which produced false-negative results for powder materials, reference spectra were acquired through the bottles listed in Figure 2 and then re-tested. Figure 3 summarizes the results of Destructive Limit Testing (DLT) to determine how many layers of bags were required to produce a false-negative result.

Figure 3

Figure 4

Conclusions

The results demonstrate that most common packaging materials make a negligible contribution to the Raman spectra of materials contained within them. The strong Raman response for solvents makes their identity easy to verify regardless of packaging. For powdered materials, the effect of plastic bags on spectral properties was minimal, and with polyethylene, accurate identification could be obtained through up to 35 layers of bags for strong Raman scattering materials such as Titanium Dioxide. The Raman response of polystyrene, which prevented positive identification based on neat spectra, was effectively compensated by acquiring additional reference spectra through polystyrene containers. Thick-walled glass proved to be the most challenging due to variations in the luminescence profile between different glass thicknesses and suppliers that interfered with the spectra of weak Raman-scattering materials, even with a reference spectrum acquired through thick glass.

Ahura Scientific, Inc.

46 Jonspin Road • Wilmington, MA 01887

Tel. (978) 642-2549, Fax (978) 657-5921

www.ahurascientific.com • truscan@ahurascientific.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.