Laser Frequency Offset Locking

Optical phase-locking is a common technique for transferring the frequency and phase characteristics of one laser beam to another. It is commonly used in heterodyne metrology, free-space optical communications, and spectroscopic applications. In this technical note we discuss the implementation of offset phase locking using a digital phasemeter and characterize the stability of phase locking two lasers.

Optical Offset Phase Locking

To put it simply, offset phase locking stabilizes phase difference between two lasers. This is done by a relatively simple process combining the outputs of the two lasers (via a beamsplitter or similar) and illuminating a photodetector with the combined beam. The combined power on the photodetector PPD can be described by:

Where ω1 and ω2 denote the frequency, and φ1 and φ2 denote the phase of each laser. Note that the high order terms are generally outside the bandwidth of the photodetectors. It is important to recognize that even though the beat note contains phase information of the lasers, this information is contained in the argument of the signal and is relatively difficult to use in feedback systems in this form. To extract the phase from the beat note we use a phase detector.

Moku:Pro Phasemeter

Although a mixer and low pass filter are adequate for most offset phase locking system, they are not without their limitations. For one, the range of the mixer-filter combination is limited to ±π/2, and the phase output from the system is linear only when very close to zero. These range and linearity issues often make it difficult to deal with systems with large fluctuations. Moku:Pro Phasemeter uses phase-locked loops (PLL) as the phase detectors, and provides automatic phase unwrap function to extend the linear dynamic range of a standard mixer-type phase detector. The PLL can assist in initial lock acquisition, removal of nonlinear affects (such as cycle slipping), and ensure a more robust lock.

Experimental Setup

We demonstrated Moku:Pro's Phasemeter offset phase locking architecture using two nonplanar ring oscillators (NPRO) lasers. The beams of the master and slave lasers were combined by a beamsplitter and interfered at a photodiode as shown in Figure 1. The beat note was connected to Moku:Pro's input 1. The feedback signal was then connected to the frequency controller of the following laser.

Lock Performance

To evaluate the phase stability of the lock, an additional Moku:Lab Phasemeter was connected to the photodetector to capture the frequency and phase drift as a function of time. Taking the amplitude spectral density of the frequency, we see that an improvement in the stability of over 4 orders of magnitude was measured, with relative frequency stability of 1 Hz/√Hz measured down to 0.1 Hz.

Questions?

support@liquidinstruments.com

Liquid Instruments

740 Lomas Santa Fe Dr Ste 102, Solana Beach, CA 92075

Tel. (619)-332-6230

Website: www.liquidinstruments.com


Related to this article

A robotic arm in an automated agricultural setting, carefully processing a conveyer belt with tomatoes and leafy greens for food production. | Image Credit: © Donnangelo - stock.adobe.com
Croatian researchers outline how optical emission spectroscopy (OES) can track reactive species and plasma parameters in real time, giving processors a path to move cold plasma treatments from the laboratory to production lines.
Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.