Computerized System Validation (CSV) is usually associated with great mountains of paper (GMP). Why should this be the case? Ask yourself, am I lazy? Why write multiple documents if only one is required to validate a spectrometer with a simple intended use? Interested? Read on…
Today’s the day you have been waiting for! Your brand-new UV-visible (UV-vis) or infrared (IR) spectrometer and controlling software is due to arrive in your regulated laboratory. The instrument’s location and services are ready. The supplier’s engineer is on hand to install and qualify the instrument against your user requirements. You’ve reviewed and approved the supplier’s qualification protocol and reports for installation qualification (IQ) and operational qualification (OQ). All goes smoothly and you are ready to evaluate your new toy, when you hear the thud of heavy footsteps behind you. The Compliance Police have arrived to ruin your day!
The Compliance Police (Quality Assurance [QA] ± CSV) have descended and you are about to undergo the third degree questioning:
You can’t use the system until all these documents have been generated and approved. What?! All this documentation for a simple spectrometer that is just going to read absorption at specified wavelengths or verify the identity of a substance by comparison with a reference standard? Consider yourself lucky. You could have been asked to write a functional specification as well!
A risk-averse CSV approach can result in writing all these documents, especially if there is a one-size-fits-all procedure that is applied regardless of the intended use of the instrument and software. This is what gets CSV a bad name–inflexible organizations who just want to play it safe.
Does it have to be this way? No.
The needs associated with the software and instrument hardware need to be addressed, and for simple intended use, this can be achieved in an integrated validation document.
To understand the approach, let’s start with regulations. We need to apply a 21 CFR 211.160(b) scientifically sound (1), flexible, common sense and logical approach to CSV. The key to doing this is understanding the 5 P’s:
Procedures: Inputs are the regulations, regulatory guidance, and applicable pharmacopoeial general chapters that result in SOPs and analytical procedures for performing work. Flexible procedures that allow an integrated approach to Analytical Instrument Qualification (AIQ) and Computerized System Validation (CSV) are mandatory. Defining the intended use of the instrument or system (1,2) is the key to determining the required documents for qualification, validation, or both.
Process: Know and understand the process to be automated. Streamline and simplify the process, implement electronic signatures and eliminate dross such as paper and spreadsheets (3–5). This allows the laboratory to define the user requirements and configuration settings to ensure data integrity. Unfortunately, process redesign is not mentioned in the draft FDA Computer Software Assurance (CSA) guidance (6).
Product: The system consists of the instrument and the software that controls it. We must select the right system to digitalize a process with appropriate technical controls for data integrity. There is always an interaction between the product and the process to fine-tune configuration settings. Purchasing the right system that can digitalize a process and ensure data integrity must take precedence over purchasing on price. When application software is involved, it is critical that the supplier’s software development is assessed to see if it can be leveraged into the laboratory validation to reduce the effort required.
People: A multi-disciplinary team approach is required between key users, laboratory managers (Data Owner/Process Owner), QA, CSV and IT (System Owner), in addition to the supplier. The team must be trained, flexible, and manage risk effectively to qualify the instrument and validate the system, especially with the amount and extent of documentation required. When it comes to people, the role of the senior management is critical to lead and support the entire 5 P’s approach.
Project: The 4 P’s above come together in a project covering the definition, selection, qualification, and validation of the system. The scope and extent of the project depends on the intended use, process to be automated, records to be created and managed by the system.
The way the 5 P’s come together for an integrated approach is shown in Figure 1.
FIGURE 1: The 5 P’s of an integrated AIQ and CSV approach.
To determine how much qualification and validation work needs to be performed, we need to time travel back to 2013, when Chris Burgess and I published a system risk assessment for analytical instruments and systems (7). One of the outcomes from the risk assessment was a sentence that referenced an article that described a single Integrated Validation Document (IVD) (8).
In essence, if the controlling software application is GAMP software category 3 and a commercially available non-configured product (9), even if it generates data used for analyses of clinical trial materials, product submissions or product releases, it could be validated using the IVD approach. Before discussing the IVD though, it is important to understand the difference between parameterization and configuration of software.
Knowing how to differentiate between GAMP software categories 3 and 4 typically involves an arm wrestling match (sorry, collaboration) between the laboratory manager and CSV. Category 3 software cannot change the business process it automates; it does exactly what it says in the brochure, nothing more, nothing less. Although run time configuration is possible (for example, definition of user roles and access privileges, report formats), the business process cannot be changed.
A major cause of confusion is parameterization, which can be performed on functions in both category 3 and 4 software. Parameterization is setting the wavelength used to measure absorption of an analyte. If the wavelength is set at 220 nm and then changed to 280 nm, the business process has not changed. The instrument is still measuring the absorbance of an analyte regardless of wavelength, meaning the business process is unchanged. However, many people think this is configuration despite it being parameterization, so it still needs to be controlled when people use the system for things like method validation.
EU GMP Annex 15 in clause 2.5 states:
Qualification documents may be combined together, where appropriate, e.g. installation qualification (IQ) and operational qualification (OQ) (10).
The principle of the IVD condenses the whole suite of validation documents into one. From the system risk assessment (7), this approach is only applicable to lower risk systems typically, but not exclusively, GAMP Category 3 software.
A description of the rational and justification can be found in the article by McDowall (8) which is available for free download (https://onlinelibrary.wiley.com/doi/10.1002/qaj.443).
As the name suggests, all key validation requirements for demonstrating intended use are contained in a single document, as shown in Figure 2. An IVD has two main sections: specifications (shown in green) and testing/reporting (shown in yellow).
FIGURE 2: Content of an integrated validation document (IVD).
The size of a typical IVD is in the range of 30–45 pages, depending on the intended use of the system.
The introduction and specifications, shown in green in Figure 2, consist of:
Not shown in Figure 2 is a referenced documents section linking an IVD with regulations, CSV procedures, and applicable SOPs for operation of the system, such as user account management. In addition, supplier assessment and qualification documentation (for example, IQ, OQ, or calibration) will also be referenced and, where possible, used to verify user requirements.
The second section of an IVD, colored yellow in Figure 2, consists of test procedures to confirm intended use, test execution notes to capture unexpected events during testing, and finishing with a reporting section with a simple release statement for operational use:
The whole document is reviewed for technical content by the laboratory and approved by QA prior to execution.
After execution by the tester, they complete the Test Summary and Release Statement section of the IVD, stating whether or not the system can be released for operational use. The whole document is reviewed by a second person, who also confirms the release of the system. QA also reviews both the completed IVD and documented evidence.
You may think that this is a new approach to integrated qualification and validation of laboratory systems; however, it has been used for nearly 20 years. There have been minor changes in approach over the intervening years, but it is essentially the same as published. Go on, be lazy. You know it’s common sense, but will the Compliance Police let you?
I would like to thank Paul Smith and Mahboubeh Lotfinia for their constructive review comments in preparation of this article.
(1) Food and Drug Administration 21 CFR 211 Current Good Manufacturing Practice for Finished Pharmaceutical Products. Silver Spring, MD, 2008.
(2) United States Pharmacopoeia Convention Inc. USP General Chapter <1058> Analytical Instrument Qualification. Rockville, MD.
(3) McDowall, R. D. Pharma 4.0 and the Digital Regulated Laboratory Part 1: Why Digitalize Your Regulated Laboratory? LCGC 2022.
(4) McDowall, R. D. Pharma 4.0 and the Digital Regulated Laboratory Part 2: Digital Laboratory Automation Strategy and Process Mapping. LCGC 2022.
(5) McDowall, R. D. Pharma 4.0 and the Digital Lab eBook 3: Quick Wins and Project Implementation. LCGC 2022.
(6) Food and Drug Administration. FDA Draft Guidance for Industry Computer Software Assurance for Production and Quality System Software. Silver Spring, MD, 2022.
(7) Burgess, C.; McDowall, R. D. An Integrated Risk Assessment for Analytical Instruments and Computerized Laboratory Systems. Spectroscopy 2013, 28 (11), 21–26.
(8) McDowall, R. D. Validation of Computerized Systems Using a Single Life Cycle Document (Integrated Validation Document). Qual. Assur. J. 2009, 12, 64–78. DOI: 10.1002/qaj.443
(9) International Society of Pharmaceutical Engineering. Good Automated Manufacturing Practice (GAMP) Guide 5, Second Edition. Tampa, FL, 2022.
(10) European Commission. EudraLex - Volume 4 Good Manufacturing Practice (GMP) Guidelines, Annex 15 Qualification and Validation. Brussels, Belgium, 2015.
(11) McDowall, R. D. Does CSA Mean “Complete Stupidity Assured?” Spectroscopy 2021, 36 (9), 15–22.
(12) European Commission. EudraLex - Volume 4 Good Manufacturing Practice (GMP) Guidelines, Chapter 6 Quality Control. Brussels, Belgium, 2014.
(13) McDowall, R. D. CSA: Much Ado About Nothing? Spectroscopy 2023, 38 (4), 7–13, 34.
R.D. McDowall is the director of R.D. McDowall Limited and the editor of the “Questions of Quality” column for LCGC Europe, Spectroscopy’s sister magazine. Direct correspondence to: SpectroscopyEdit@MMHGroup.com●
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.
An Enhanced Approach to Analytical Instrument and System Qualification
May 1st 2025United States Pharmacopoeia (USP) general chapter <1058> on Analytical Instrument Qualification (AIQ) now has a new title from a draft update for public comment: Analytical Instrument and System Qualification (AISQ). A new three-phase integrated lifecycle phase approach to qualification and validation is described. But will laboratories and suppliers implement it?
Synthesizing Synthetic Oligonucleotides: An Interview with the CEO of Oligo Factory
February 6th 2024LCGC and Spectroscopy Editor Patrick Lavery spoke with Oligo Factory CEO Chris Boggess about the company’s recently attained compliance with Good Manufacturing Practice (GMP) International Conference on Harmonisation of Technical Requirements for Registration of Pharmaceuticals for Human Use (ICH) Expert Working Group (Q7) guidance and its distinction from Research Use Only (RUO) and International Organization for Standardization (ISO) 13485 designations.