|Articles|February 1, 2016

Application Notebook

  • Application Notebook-02-01-2016
  • Issue 0

Single Reaction Chamber Microwave Digestion for Pharmaceutical Samples in Accordance with USP Chapters and

The United States Pharmacopeia (USP) is in the process of developing new USP chapters and for the measurement of inorganic contaminants in pharmaceutical samples. While samples can be dissolved in aqueous or organic solvents, the large majority will require digestion, and closed vessel microwave digestion is stipulated by USP. Single reaction chamber (SRC) microwave digestion is a new type of closed vessel digestion that was developed by Milestone and offers many advantages over closed vessel digestion-particularly for pharmaceutical sample types.

Single reaction chamber microwave digestion offers significant benefits over traditional microwave digestion for pharmaceutical samples: higher sample throughput, reduced labor and consumables costs, simplified workflow, and superior digestion quality.

The United States Pharmacopeia (USP) is in the process of developing new USP chapters <232> and <233> for the measurement of inorganic contaminants in pharmaceutical samples. While samples can be dissolved in aqueous or organic solvents, the large majority will require digestion, and closed vessel microwave digestion is stipulated by USP. Single reaction chamber (SRC) microwave digestion is a new type of closed vessel digestion that was developed by Milestone and offers many advantages over closed vessel digestion-particularly for pharmaceutical sample types.

Instrumentation

The Milestone UltraWAVE can digest up to 15 samples simultaneously, at up to 300 °C and 199 bar pressure. The high temperature and pressure capabilities enable the complete digestion of virtually every sample type. Samples are weighed into vials and placed in the chamber: any combination of sample types can be digested together-greatly simplifying the sample prep workflow. Raw materials, excipients, API, and final product can all be digested together in the same run. The Milestone UltraWAVE uses disposable glass vials, eliminating vessel assembly/disassembly and vessel cleaning, which significantly reduces labor and consumables costs. Most pharmaceutical sample types have high organic content, including oils and fats that are difficult to digest and generate CO2, which increases pressure in the reaction vessel. The high pressure capability of the UltraWAVE enables it to digest entire fish oil gelcaps-not possible with traditional closed vessel microwave.


Table I: St. John's wort

The UltraWAVE can digest samples in as little as 3 mL acid, which minimizes reagent blank levels-important since ICP-MS will be used predominantly for USP <232> and <233> when implemented.


Table II: Fish oil gelcaps

The SRC chamber is pre-pressurized with N2 prior to the start of the run. This prevents sample boiling, eliminating cross-contamination and loss of volatiles.


Table III: Quality control summary (fish oil gelcaps)

Due to the UltraWAVE's higher sample capacity, use of disposable vials, and faster cool down time, sample throughput is 2–3× higher than closed vessel digestion. Its lower consumables costs, simple operation, and superior digestion quality also make the Milestone UltraWAVE the perfect choice for pharmaceutical sample prep.

Results

The data above shows excellent DLs and recovery-even for the volatile element mercury-no cross-contamination with the UltraWAVE.

Milestone, Inc.
25 Controls Drive, Shelton, CT 06484
tel. (866) 995-5100, fax (203) 925-4241
Website: www.milestonesci.com


Related to this article

Atoms Don't Lie–The Remaking of Atomic Spectroscopy, © Hallie -chronicles-stock.adobe.com
Atomic spectroscopy has quietly gone from counting elements in bulk to counting single nanoparticles, imaging tissue at the micrometer scale, and reading Martian rocks from several meters away. The instruments got smaller, the software got smarter, and the questions they can now answer got a lot more interesting.
PFAS spiking experiments in bean plants (Phaseolus vulgaris)
Atomic Perspectives
This Atomic Perspectives column explains how high-resolution continuum source graphite furnace molecular absorption spectrometry (HR-CS-GFMAS) quantifies extractable organic fluorine (EOF) as a PFAS sum parameter, complementing LC-MS/MS target analysis, with applications to German river sediments, bean plants, and fluorinated polymer sealing materials.