Trace-Level Pesticide Detection Utilizing Surface-Enhanced Raman Spectroscopy

Author(s)Ocean Insight

Several pesticides were detected at parts-per-million levels on apple skin using a unique swabbing method and surface-enhanced Raman spectroscopy (SERS).

Commonly used pesticides, such as organothiophosphates and fungicides, can attack the central nervous system, posing a risk to humans and other animals upon exposure (1). SERS has been investigated as a technique to detect trace levels of pesticides on food items, and offers reduced instrument size, fast measurement times, non-destructive sampling, and simple implementation compared to the traditionally used methods, such as HPLC and GC–MS (2). In this study, we utilize a swabbing technique that has the potential to be a testing platform for field use to detect trace levels of pesticides on apple skin.

Experimental Conditions

Pesticide-containing apple skin pieces are swabbed and added to colloidal gold nanoparticles. The mixtures are interrogated with 785 nm excitation laser with 3 s integration time and 350 mW laser power, and the Ocean Insight QE Pro spectrometer.

Results

The pesticides examined in this study were thiram, malathion, acetamiprid, and phosmet. The SERS spectra of each pesticide after being applied to apple skin and swabbed are presented in Figure 1. The observed peaks are consistent with reports in the literature.


Figure 1: SERS spectra of swabs of 1 ppm (a) malathion, (b) acetamiprid, (c) phosmet, and (d) thiram on apple skin.

Conclusions

We present a swabbing technique that has the potential to be a testing platform for field use and utilizes colloidal gold to detect trace levels of several pesticides on apple skin. This technique can detect each pesticide down to 1 ppm, where the pesticide residue tolerances on apples are 5 ppm, 8 ppm, 1 ppm, and 10 ppm for thiram, malathion, acetamiprid, and phosmet, respectively (3). The results presented here indicate that SERS coupled with the swab method is a valuable tool, and has significant potential for identifying pesticide residues on the surface of fruits for food quality and safety control.

References

(1) M. Stoytcheva, InTech, 30–48 (2011). doi: 10.5772/1003.

(2) M.L. Xu, Y. Gao, X.X. Han, and B. Zhao, J. Agric. Food Chem. 65, 6719–6726 (2017).

(3) Office of the Federal Register, Electronic Code of Federal Regulations, 2018, https://www.ecfr.gov (8 January 2019).

Ocean Insight
8060 Bryan Dairy Road, Largo, FL 33777
tel. +1 (727) 733-2447
Website: www.oceaninsight.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.
Highlights from What’s Nu October 2026
In this Q&A, we recap the October edition of “What’s Nu,” which covered topics such as how novel spectroscopic and analytical methods are transforming environmental monitoring, medical treatment, and accessibility in scientific research.