Milestone's UltraWAVE microwave digestion system with its groundbreaking single reaction chamber (SRC) technology enables the simultaneous digestion of up of to 22 different food sample types with different chemistries at temperatures and pressures as high as 300 °C and 200 bar, greatly increasing efficiency and simplifying product development.
Elemental impurities are monitored in foods to ensure safe levels and compliance with regulation. One particular class of these elemental impurities, the "Big 4" (arsenic, mercury, lead and cadmium), are of special concern because ingestion of quantity relatively low in food product can cause serious health issues, including - but not limited to - reproductive troubles, developmental disorder, brain damage, kidney and liver damage, and cancer. Given the serious health risks, a more stringent regulation is calling for the use of closed vessels digestion for sample preparation and inductively coupled plasma (ICP) for elemental impurities analysis as a replacement of obsolete and subjective methods of analysis based on colorimetric tests.
In this application note, two frosting samples are digested simultaneously on a 15-position rack with a single temperature profile. In addition, setting the instrument is straightforward: a ramp time to target temperature and a hold time at target temperature.
Frosting or icing products are delicate matrices to digest because they are made of significant amounts of carbohydrate and lipids, and thus are very exothermic during degradation. Digestion of the two varieties of frosting using a traditional microwave system would require two different runs and could possibly require two different rotors. Because of the exothermic nature of the sample, on a traditional microwave there is a limitation of the size of the sample that could be digested, since a relatively large sample size (>0.500 g) will generate a level of pressure that will cause the system to vent with loss of volatile elements.
Table I: Sample and chemistry
Samples were placed in a 15-position rack and lowered into the single reaction chamber. The chamber was pre-pressurized with nitrogen at 40 bar to raise the acids boiling point and thus prevent any cross-contamination and loss of volatiles due to boiling and spitting as the temperature increases inside the chamber.
The two samples were fully digested as evidenced by the clear solutions obtained. The digestion lasted for less than 1 h including cooling time. The higher performance and greater ease of use make the UltraWAVE a good fit for a food company where samples of different types are common. Additionally, by using five position vials, samples as large as 2 g can be digested. Such capability not only addresses the common sample homogeneity concern of the food industry, but also gives the possibility to dilute samples with as much flexibility for ICP-OES as for ICP-MS.
Milestone, Inc.
25 Controls Drive, Shelton, CT 06484
tel. (866) 995-5100, fax (203) 925-4241
Website: www.milestonesci.com
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.
Specificity and the Net Analyte Signal in Full-Spectrum Analysis
July 21st 2025This tutorial addresses the critical issue of analyte specificity in multivariate spectroscopy using the concept of Net Analyte Signal (NAS). NAS allows chemometricians to isolate the portion of the signal that is unique to the analyte of interest, thereby enhancing model interpretability and robustness in the presence of interfering species. While this tutorial introduces the foundational concepts for beginners, it also includes selected advanced topics to bridge toward expert-level applications and future research. The tutorial covers the mathematical foundation of NAS, its application in regression models like partial least squares (PLS), and emerging methods to optimize specificity and variable selection. Applications in pharmaceuticals, clinical diagnostics, and industrial process control are also discussed.
New Study Expands Nickel Autoionization Spectra to Advance Laser Isotope Separation Technologies
July 17th 2025Researchers at China’s National Key Laboratory have identified 170 nickel autoionization states using resonance ionization mass spectrometry, significantly advancing the spectral database critical for laser isotope separation and atomic spectroscopy.
AI-Powered Fusion Model Improves Detection of Microplastics in the Atmosphere
July 17th 2025Researchers from Nanjing University of Information Science & Technology have introduced a breakthrough AI-enhanced multimodal strategy for real-time detection of polyamide microplastics contaminated with heavy metals.
How Analytical Chemists Are Navigating DOGE-Driven Funding Cuts
July 14th 2025DOGE-related federal funding cuts have sharply reduced salaries, lab budgets, and graduate support in academia. Researchers view the politically driven shifts in priorities as part of recurring systemic issues in U.S. science funding during administrative transitions. The impact on Federal laboratories has varied, with some seeing immediate effects and others experiencing more gradual effects. In general, there is rising uncertainty over future appropriations. Sustainable recovery may require structural reforms, leaner administration, and stronger industry-academia collaboration. New commentary underscores similar challenges, noting scaled-back graduate admissions, spending freezes, and a pervasive sense of overwhelming stress among faculty, students, and staff. This article addresses these issues for the analytical chemistry community.