Glass Analysis
The standard procedure for the determination of the emissivity of glass involves single beam measurements of the glass sample, a reference mirror, and an empty background spectrum at 30 standard wavelengths. These wavelengths range from 5.5 μm to 50 μm (1818–200 cm-1). The selected wavelengths are specified by three international standards: EN 673 (1), EN 12898 (2), and ISO 10292 (3). The ability of the PerkinElmer Spectrum 3TM (Figure 1) to measure in multiple regions of the infrared spectrum gives the user the potential to not only determine the emissivity of glass, but investigate other properties demonstrated in the near-infrared region.
Figure 1: PerkinElmer Spectrum 3 FT-IR spectrometer.
Experimental
A 3 mm-thick sample of float glass was measured using the PerkinElmer Spectrum 3 FT-IR spectrometer with the OMT Specular Reflectance set. For each of the glass, reference mirror, and empty background, three measurements were made. The three spectral ranges used are shown in Table I.
The Spectrum 3 has the ability to scan the entire infrared range from near-infrared to far-infrared with automatic switching of optical components. The new “Merge” function in Spectrum 10™ may be used to combine spectra from different ranges to produce a single spectrum which covers the entire infrared region.
Results and Discussion
The merged spectra of the glass sample, the reference mirror, and the background spectrum are shown in Figure 2.
Figure 2: Tri-range spectra of glass (Red), a reference mirror (Blue) and no sample (Black).
The emissivity of glass is determined from the reflectance at room temperature using equation 1.
εn = 1 – Rn [1]
where εn is the emissivity at normal incidence, and Rn is the reflectance at normal incidence. The reflectance is calculated as an average of 30 spectral reflectance values measured at specified wavelengths between 5 and 50 μm (1818.8–200.0 cm-1).
First, both the glass and mirror measurements are corrected using the empty background spectrum. The reflectance is then determined at each wavelength by dividing the corrected glass value by the corrected mirror value. The overall reflectance is calculated by finding the mean of the reflectance at each of the 30 specified wavelengths. This may be carried out using the “equations” function of Spectrum 10 software or with an external data analysis application.
Using this calculation, the emissivity of the 3 mm float glass sample was found to be approximately 0.95, in agreement with the commonly accepted values for glass emissivity at room temperature (4).
In addition to emissivity calculations, the near-infrared region of the spectrum can also provide information regarding thermal transmittance. Glass designed specifically for absorbing heat will generally have a higher reflectance in the near-infrared region. For example, the standard 3 mm float glass shows low reflectance in the near-infrared region, and thus is not heat-absorbing glass.
Pharmaceutical Analysis
A pharmaceutical product which demonstrates strong spectral features in all three regions is clotrimazole, an anti-fungal medication, the chemical structure of which is shown in Figure 3.
Figure 3: Chemical structure of clotrimazole.
Experimental
A sample of clotrimazole was measured in all three regions of the infrared spectrum, using automatic switching of optical components. Measurements in the mid and far-infrared regions were carried out using an ATR accessory, while measurements in the near-infrared region were carried out using the NIRAII accessory.
Results and Discussion
The tri-range spectrum of clotrimazole is shown in Figure 4.
Figure 4: Tri-range spectrum of clotrimazole.
Having the ability to measure a pharmaceutical sample in all three ranges of the infrared region through Spectrum 3 allows analysts to simultaneously investigate multiple properties and obtain more significant information on the sample.
Industrial Antioxidant Analysis
One example of a material used across a wide range of industries is butylated hydroxytoluene (BHT). This is used as an antioxidant in transformer oil, fuels, rubber, pharmaceuticals, and some food products. The chemical structure of BHT is shown in Figure 5.
Figure 5: Chemical structure of BHT.
Experimental
The PerkinElmer Spectrum was used to measure a sample of BHT in all three regions of the infrared spectrum. Measurements in the mid and far-infrared regions were carried out using an ATR accessory, while measurements in the near-infrared region were carried out using the NIRAII accessory.
Results
A tri-range spectrum of BHT is shown in Figure 6.
Figure 6: Tri-range spectrum of BHT.
As with other materials, different areas of the infrared spectrum of BHT serve different purposes. For example, the sharp peak at around 3700 cm-1 has been used for quantitative determination of BHT in food samples such as edible oils (5).
One previously mentioned use of BHT is as an antioxidant in rubber and other polymers. Far-infrared provides an incredibly useful tool for measurement of polymers due to strong intermolecular vibrations and weak intramolecular vibrations falling in this range. This region has been used to successfully quantify BHT as an additive in polymer samples as lattice vibrations of some polymers fall between 50–100 cm-1 (6).
References
PerkinElmer, Inc.
940 Winter Street, Waltham, MA 02451
Tel. (781) 663-6900
Website: www.perkinelmer.com
Get essential updates on the latest spectroscopy technologies, regulatory standards, and best practices—subscribe today to Spectroscopy.
New Raman Probe Shows Promise for Real-Time Soft Tissue Sarcoma Detection During Surgery
July 30th 2025Researchers from Université de Montréal have demonstrated that a new handheld Raman spectroscopy device, the UltraProbe, can accurately and rapidly detect retroperitoneal soft tissue sarcomas during surgery, offering real-time tissue analysis that could improve surgical outcomes.
Best of the Week: The Emerging Leader in Molecular Spectroscopy, Big Pharma’s Manufacturing Shift
July 25th 2025Top articles published this week include a feature article about big pharma’s investments in U.S.-based manufacturing, an article about the 2025 Emerging Leader in Molecular Spectroscopy Lingyan Shi, and some news items detailing the winners of the Coblentz Society’s student awards.