Common Problems with FT-IR Instruments and How to Avoid Them

Listen
0:00 / 0:00

Key Takeaways

  • Lack of formal training in FT-IR spectroscopy can lead to difficulties in identifying quality spectra, causing data processing issues.
  • Instrument malfunctions and environmental vibrations can introduce unwanted spectral features, affecting data accuracy.
SHOW MORE

Users of Fourier transform infrared (FT-IR) spectrometers may have received little or no formal training in spectroscopy. Therefore, some users cannot distinguish between “good” and “bad” spectra, which can lead to problems in spectral data processing. Many of the problems that are commonly encountered with the FT-IR spectra measured are easy to avoid if one is aware of the problem. The sources of the problem can be loosely categorized into instrument, accessory, sample, and data processing.

To thoroughly understand the operation of a FT-IR instrument, one needs to realize that the instrument inherently generates a data set that undergoes mathematical transformations, and instrument malfunctions can manifest themselves as spectral features after the transformation. A good example is features that might show up in a spectrum because of environmental vibrations, such as bumping an instrument or having a vacuum pump running on the same bench. By compiling a background with an empty beam and (no accessory or sample) then collecting a sample under the same conditions and ratioing in transmission, this problem can be seen. For a good discussion of the instrument theory and effects, see the literature (1).

Attenuated total reflection (ATR) is one of the most common and easiest sampling techniques. However, ATR is a technique that can generate false data. ATR interrogates the surface of the sample so that the chemistries on the surface are amplified. One of the most common problems encountered in an FT-IR–ATR analysis is taking the background spectrum with a dirty ATR element. In the spectrum shown in Figure 1a, we see negative features entering the absorbance spectrum, which indicates that the ATR element was not clean when the background was collected. A quick wipe of the ATR element and a new background collection give the actual spectrum of the sample in Figure 1b.

To illustrate the problems that may result from the sample, we can again look at an ATR analysis. Many plastic materials may show different spectra on the surface and slightly into the bulk. Plasticizers can migrate to or away from the surface; the surface may also be oxidized while the bulk represents the actual sample, or there may be some other surface effect because of processing. Figure 1c illustrates this problem. The red spectrum shows the surface of the sample run as the sample was received. The C–H stretch region is fairly weak and the ratios of the peaks on the band envelope at around 1100 cm-1 show different ratios. The blue spectrum on the top shows the sample after the outer surface has been cut away and is more representative of the bulk of the sample.

This surface effect that may appear to be a problem can be used advantageously if you suspect some surface effect either by collecting a spectrum and then cutting into the sample, or making use of the relationship of depth of penetration to the angle of incidence and refractive index of the ATR element. By varying these combinations without violating the critical angle, you can achieve different penetration depths. After an ATR correction to account for the wavelength dependent penetration depth, you can compare surface chemistries. There are two good references to the theory and application of the technique, one by Harrick and the other by Milosevic (2,3).

Finally, as mentioned above, the wrong data processing technique can distort the spectrum. A good example is shown in Figure 2. A spectrum collected in diffuse reflection should be ratioed into Kubelka-Munk (K-M) units. In the literature, there is a book chapter written by Griffiths and de Haseth on the topic (4). When the spectrum is ratioed in absorbance (top), the peaks are distorted and appear saturated, and minimal spectral information is available. By calculating the ratio in K-M units, the spectrum now looks normal and can be interpreted.

As illustrated, there are many sources of problems that can appear. If you understand the implications of the instrument, the accessory, the sample, and the data processing on the final result, you will get a good answer to the analysis question.

References

  1. P.R. Griffiths and J.A. de Haseth, Fourier Transform Infrared Spectrometry (Wiley & Sons, Inc., Hoboken, New Jersey, 2006).
  2. N.J. Harrick, Internal Reflection Spectroscopy, 1st ed. (Wiley Interscience, Hoboken, New Jersey, 1967), pp. 327.
  3. M. Milosevic, Internal Reflection and ATR Spectroscopy (Wiley & Sons, Inc., Hoboken, New Jersey, 2012).
  4. P.R. Griffiths and J.A. de Haseth, in Fourier Transform Infrared Spectrometry (Wiley & Sons, Inc., Hoboken, New Jersey, 2006), pp. 349–362.

Related to this article

Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.
Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Highlights from What’s Nu October 2026
In this Q&A, we recap the October edition of “What’s Nu,” which covered topics such as how novel spectroscopic and analytical methods are transforming environmental monitoring, medical treatment, and accessibility in scientific research.