Eliminating the Trade-Off Between Resolution and Throughput in Raman Spectroscopy

Although Raman spectroscopy is an excellent analytical tool, Raman signals are often weak and traditional slit spectrometers typically have poor optical throughput, limiting their effectiveness in low light level setups.

Although Raman spectroscopy is an excellent analytical tool, Raman signals are often weak and traditional slit spectrometers typically have poor optical throughput, limiting their effectiveness in low light level setups. Using a high throughput virtual slit design, the Apex 785 Raman spectrometer from Ocean Optics eliminates trade-offs between sensitivity and resolution.

Conventional spectrometers use a narrow entrance slit to achieve high resolution, but at the expense of throughput. The slit can reject as much as 75–95% of the light entering the optical bench. With high-throughput virtual slit (HTVS) technology licensed from Tornado Spectral Systems, Apex spectrometers overcome the compromise between throughput and spectral resolution.

Apex uses a beam reformatting technology that eliminates the need for a physical slit to achieve high resolution. When combined with precision optical coatings, volume phase holographic (VPH) gratings, and solid state detector technology, spectrometer throughput as high as 95% is achievable.

Also, Apex spectrometers use a large input aperture and specially configured optical components to compress, reformat, and then expand the light beam, which narrows the input aperture along the dispersion axis while preserving total flux.

Experimental Conditions

To maintain peak optical throughput, Apex uses a 1200 line/mm VPH grating blazed at 830 nm in an optical bench with an uncooled, NIR enhanced back-thinned CCD array detector.

Raman measurements of paracetamol (acetaminophen), polystyrene, isopropanol, and toluene were made using the Apex spectrometer (APEX-RAMAN-785) in a setup with a 100 mW 785 nm laser diode, a Raman sample holder and a 785 nm Raman fiber optic probe. Integration time intervals ranged from

195 ms for toluene to 750 ms for paracetamol.

Results

As shown in Figure 1, the Apex measures spectral data over a Raman shift range beyond 3800 cm-1 with high spectral density, delivering Raman data with more spectral features for better library matching and sample identification. Also, the Apex spectrometer enables the acquisition of high quality Raman spectral data at very short integration times.

Figure 1: This Raman spectrum for paracetamol was acquired at 750 ms integration time.

Typical Raman measurements require seconds to achieve similar signal levels. This rapid measurement speed enables measurements in process flows and for samples that change over time. The low integration times required for the Apex also allow for the measurement of multiple samples in the time it takes to obtain one spectrum from a less sensitive spectrometer. This provides the opportunity for signal averaging to increase SNR, further improving measurement accuracy.

Conclusions

A new spectrometer design based on HTVS optical slicer technology has been shown to provide a significant increase (a factor of 10–15×) in optical throughput without any detriment in optical resolution. In addition, the technology allows users to achieve high sensitivity without the need for long integration times, detector cooling, or high powered excitation sources; to use shorter measurement times for faster measurements; to acquire more accurate spectral signatures; and to use lower power excitation sources to minimize photodegradation.

References

(1) Tornado Spectral Systems contributed to this application note.

Ocean Optics

830 Douglas Avenue, Dunedin, FL 34698

tel. (727) 733-2447, fax (727) 733-3962

Website: www.OceanOptics.com


Related to this article

San Francisco skyline with Oakland Bay Bridge at sunset, California, USA | Image Credit: © JFL Photography - stock.adobe.com
The January 30–February 4 event is expected to feature more than 4,500 presentations across six technical areas and more than 1,400 exhibitors at the Moscone Center.
A robotic arm in an automated agricultural setting, carefully processing a conveyer belt with tomatoes and leafy greens for food production. | Image Credit: © Donnangelo - stock.adobe.com
Croatian researchers outline how optical emission spectroscopy (OES) can track reactive species and plasma parameters in real time, giving processors a path to move cold plasma treatments from the laboratory to production lines.
Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.