Rapid Screening in Microbiology by FT-IR

Yeast suspensions were dried and analyzed via FT-IR using an automated transmission sampling accessory. Up to six-fold sampling capacity increase may be realized through automation.

Yeast suspensions were dried and analyzed via FT-IR using an automated transmission sampling accessory. Up to six-fold sampling capacity increase may be realized through automation.

FT-IR spectroscopy has been shown to be an effective tool to classify bacteria, yeast, and fungi at the strain, species, and genus level. Compared to traditional microbiological techniques, IR sampling is less time-consuming. Microbiological applications for FT-IR analysis include a rapid method of identification and screening, quality control for food and pharmaceutical fields, and assessment of growth environment on cell and cell byproduct physiology.

An IR plate-reader for biological samples is desirable for laboratories confronted with high-throughput demands, and the need to save resources through automation. This application note illustrates the use of an in-compartment XY plate reader to analyze yeast and discusses the resource savings involved through accessory automation.

Experimental Conditions

Yeast suspensions (30 µL) were pipetted onto a sample well of a 96-well reusable Si plate, and dried at 45 °C for 1 h for subsequent infrared transmission measurements. The 96-well Si plate was loaded into the PIKE Technologies XY Autosampler Accessory, an FT-IR sample compartment automated transmission/diffuse reflection accessory. The spectral range spanned from 4000 to 600 cm-1 using a resolution of 4 cm-1 . Data collection time was 1 min. For transmission measurements the accessory's integrated DTGS detector is used to maximize the signal-to-noise ratio. Automated plate movement to each sample position is controlled by PIKE AutoPRO software, which also triggers data collection via the FT-IR spectrometer.

Results

Figure 1 shows a typical transmission spectrum of the dried yeast suspension. Some main spectral features include the C-H stretching vibrations from fatty acids between 3100–2800 cm-1 , the amide I and amide II bands between 1800–1500 cm-1 , and C-O-C and C-O-P stretching vibrations of polysaccharides between 1200–900 cm-1 . A comprehensive list of absorption band assignments may be found in Naumann (1). Application of multivariate analysis allows differentiation between unique biological materials (2,3).

Figure 1: Infrared transmission spectrum of dried yeast suspension.

Through automation the number of IR samples analyzed per day may be greatly enhanced and potential sample and data entry errors decreased. Due to variability in biological materials, replicate sample measurements are often required. If four replicate measurements are made for one sample, 84 samples may be analyzed in a typical eight-hour work day. This equates to over 330 individual spectra collected without end user input, except to load the 96-well plates once the process is started by the AutoPRO software.

In addition to biological applications, an automated XY plate reader may be used for polymer solutions and solvent residuals analysis.

Conclusions

Infrared transmission spectra of microbiological suspensions may be collected by drying the suspension on an IR transparent plate. The use of an in-compartment XY plate reader accessory provides means for automated measurements via transmission or reflectance to increase laboratory capacity.

References

(1) D. Naumann, Encyclopedia of Analytical Chemistry, R.A. Myers, Ed. 102–131 (2000).

(2) D. Naumann, Mikrochim Acta. 373–377 (1988).

(3) G. Fischer et al., J. Microbio.l Methods 63–77 (2006).

PIKE Technologies

6125 Cottonwood Drive, Madison, WI 53719

tel. (608) 274-2721, fax (608) 274-0103

Website: www.piketech.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.