Tech Tip: Choosing Raman Excitation Wavelengths

Raman spectroscopy offers a number of benefits for testing and characterization, with the selection of system components and measurement parameters offering great flexibility.

Raman spectroscopy offers a number of benefits for testing and characterization, with the selection of system components and measurement parameters offering great flexibility. In this technical tip, we’ll focus on the choice of Raman laser excitation wavelength.

 

Raman scattering requires a very high intensity light source or long spectral acquisition times in order to generate enough photons for detection. The source should also be monochromatic to interrogate the narrowest Raman features and with the greatest resolution. Since the amount of Raman scattering is proportional to the intensity of the light source, lasers are a natural choice, particularly as they can be focused very precisely into a small sample or through a window.

Raman signal is proportional to 1/λ4 of the excitation wavelength. While this makes it tempting to use a short wavelength laser, autofluorescence generated in the sample can easily degrade signal-to-noise and make Raman peaks difficult to resolve. Organic samples are particularly prone to autofluorescence, driving excitation to red or NIR wavelengths (660–830 nm). The effect is much less pronounced for inorganic materials, including carbon nanotubes and fullerenes, more commonly studied with 532 nm laser excitation.

Figure 1: Modular Raman systems begin with the choice of excitation laser and offer flexibility in measuring different materials.

 

Excitation at 1064 nm virtually eliminates autofluorescence, but also reduces the Raman signal significantly, further complicated by the reduced sensitivity of NIR spectrometers. UV excitation is also tempting, as the full Raman fingerprint region can be collected prior to the onset of autofluorescence at ~330 nm, but again technical limitations prevail-UV lasers and filters are more expensive and lower performance, while the high energy of UV light can damage samples quickly during measurement.

 

Raman Excitation Laser Options

Raman spectroscopy requires laser sources with spectral purity and wavelength stability. Our Raman lasers use an innovative stabilization technology that provides extremely narrow wavelengths that remain locked at the desired excitation wavelength regardless of ambient temperature changes. These high-power excitation sources combine integrated laser drivers and thermoelectric coolers in a compact footprint. Lasers are available for various excitation wavelengths and can be used as turnkey (with built-in laser drivers and thermoelectric cooling) or integrated (designed for use in OEM systems) packages.

Ocean Optics

830 Douglas Avenue, Dunedin, FL 34698
tel. (727) 733-2447, fax (727) 733-3962

Website: www.OceanOptics.com


Related to this article

Ingots of pure copper or pink gold on a black background. | Image Credit: Phawat - stock.adobe.com
At SciX, Kevin Wilkinson of the Universite de Montreal explains how faster data acquisition and new software have cut analysis time and how a particle’s chemical composition can link arsenic-bearing emissions back to a copper refinery’s processes.
Spectral Frontiers: Seeing the Unseeable Microspectroscopy
Microspectroscopy, the marriage of spectroscopic chemical identification with microscopic spatial resolution, has advanced from Perkin-Elmer's first commercial infrared microspectrometer in 1953 to today's nanoscale AFM-IR probes, synchrotron-brightness infrared beamlines, and compact hyperspectral tissue imaging devices being developed for AI-assisted cancer detection.
Richard Crocombe
The New York / New Jersey Section of the Society for Applied Spectroscopy will present its 2026 Gold Medal Award to Richard Crocombe at the Eastern Analytical Symposium on November 17, recognizing two decades of work that helped carry spectroscopy out of the laboratory and into the field.