This application note describes the capability of the Teledyne Leeman Lab's Prodigy7 High-Dispersion ICP for performing analysis according to SW-846 Method 6010C.
This application note describes the capability of the Teledyne Leeman Lab's Prodigy7 High-Dispersion ICP for performing analysis according to SW-846 Method 6010C. This method is applicable to groundwaters, Toxicity Characteristic Leaching Procedure (TCLP) and Extraction Procedure (EP), industrial organic wastes, soils, sludge, and sediments.
As with other US EPA mandated methods, Method 6010C requires adherence to specific procedures to ensure analytical quality control.
A Prodigy7 inductively coupled plasma (ICP) spectrometer equipped with a radial view torch and a 240-position CETAC ASX-520 autosampler was used to generate the data.
The Prodigy7 is a compact benchtop simultaneous ICP-OES system featuring a 500 mm focal length Echelle Optical System coupled with a megapixel CMOS Detector. At 28 × 28 mm, the active area of the CMOS Detector is significantly larger than any other solid-state detector currently used for ICP-OES. This combination allows Prodigy7 to achieve higher optical resolution than other solid-state detector based ICP systems. The detector also provides continuous wavelength coverage from 165 to 1100 nm permitting measurement over the entire ICP spectrum in a single reading without sacrificing wavelength range or resolution. This detector design is inherently anti-blooming and is capable of random access, nondestructive readout that results in a dynamic range of more than six orders of magnitude.
The sample introduction system consists of:
The volume of cyclonic spray chamber is low to allow for fast washout between samples, while its knockout tube or baffle, efficiently reduces the amount of sample aerosol reaching the torch.
The torch and sample introduction system are uniquely integrated into the optical system through Prodigy7's innovative Image Stabilization System, which treats the torch as an optical component by rigidly attaching it to the spectrometer.
The initial performance parameters measured for this application note were determined as specified in Method 6010C. All samples were analyzed using the Prodigy7 equipped with a radially viewed torch.
Method 6010C requires that the upper limit of the linear dynamic range (LDR) be established. This is accomplished by analyzing increasingly higher concentration analyte standards. The LDR is considered exceeded when the measured concentration is more than 10% below the true concentration. The upper limit of the LDR must be established for each wavelength utilized by determining the signal responses from a minimum of three different concentration standards across the range. The ranges that may be used for the analysis of samples should be judged by the analyst from the resulting data. A standard at the upper limit must be prepared, analyzed, and quantitated against the normal calibration curve. The calculated value must be within 10% (±10%) of the true value.
The types of samples analyzed under the scope of Method 6010C may exhibit considerable spectral complexity. As a result, interferences due to spectral overlaps may be observed. When present and uncorrected, these interferences will produce biased determinations. For this study, checks are performed for the major matrix elements, Al, Ca, Fe, Mg, Si, and P on the other analyte elements. This is done using single element solutions. Once the interferences are identified, the Prodigy7 will automatically calculate the interfering element correction factors.
A certified water sample, ERA WatR™ Pollution Trace Metals, Lot P213-500, and four certified soil samples were prepared and analyzed: ERA Metals in SewageSludG™ , Lot. D078-160 (Environmental Resource Associates, Arvada, Colorado) and Standard Reference Materials (SRM) 2709, 2710, 2710 (NIST, Gaithersburg, Maryland).
The Prodigy7 ICP readily accommodated all of the soil samples for analysis. Examination of the data indicates the results for all elements lie well within the performance acceptance limits (PAL) for the ERA soil or the range specified for the NIST SRM soils. The results demonstrate excellent precision, as indicated by the low relative standard deviation (RSD) data, indicating the Prodigy7 is operating well above the detection limit (DL) and limit of quantitation (LOQ) for the elements determined.
For the full application note visit: http://web.teledyneleemanlabs.com/spectroscopyAN1305.
Teledyne Leeman Labs
6 Wentworth Drive , Hudson, NH 03051
tel. (603) 886-8400, fax 603-886-9141
Website: www.teledyneleemanlabs.com
Testing Solutions for Metals and PFAS in Water
January 22nd 2025When it comes to water analysis, it can be challenging for labs to keep up with ever-changing testing regulations while also executing time-efficient, accurate, and risk-mitigating workflows. To ensure the safety of our water, there are a host of national and international regulators such as the US Environmental Protection Agency (EPA), World Health Organization (WHO), and the European Union (EU) that demand stringent testing methods for drinking water and wastewater. Those methods often call for fast implementation and lengthy processes, as well as high sensitivity and reliable instrumentation. This paper explains how your ICP-MS, ICP-OES, and LC-MS-MS workflows can be optimized for compliance with the latest requirements for water testing set by regulations like US EPA methods 200.8, 6010, 6020, and 537.1, along with ISO 17294-2. It will discuss the challenges faced by regulatory labs to meet requirements and present field-proven tips and tricks for simplified implementation and maximized uptime.
Practical Autodilution for ICP-MS and ICP-OES
January 20th 2025Gain insights into improving efficiency and accuracy in elemental analysis through automated dilution technology. Learn about the key capabilities of the Agilent ADS 2 system and its seamless integration with ICP-MS and ICP-OES workflows.
UV-Vis Spectroscopy: Exporting Your Measurement Out of the Instrument
January 20th 2025Optical fibers in ultraviolet-visible (UV-Vis) spectroscopy can enable measurements outside the traditional sample compartment. This paper details the components needed for fiber optic systems, such as couplers and probes, and reviews the performance of Agilent's Cary series instruments. It is crucial to choose the right fiber optic setup for a specific lab’s needs to ensure accurate and efficient measurements.