Register Now: Advances in Spectroscopy for Pharma and Biopharmaceutical Analysis Symposium
|Articles|September 1, 2013

Application Notebook

  • Application Notebook-09-01-2013
  • Volume 28
  • Issue 9

Analysis of Potable and Waste Waters by US EPA Method 200.7 Using the Optima 8300 ICP-OES and prepFAST Auto-Dilution/Calibration System

This application brief will give a summary of a more thorough application note (1), which describes the analytical capabilities of Optima? 8300 ICP-OES utilizing low flow conditions, coupled to the prepFAST? in-line, auto-dilution/calibration sample delivery system to determine a suite of environmentally-significant elements in potable and waste waters according to US EPA Method 200.7.

This application brief will give a summary of a more thorough application note (1), which describes the analytical capabilities of Optima® 8300 ICP-OES utilizing low flow conditions, coupled to the prepFAST™ in-line, auto-dilution/calibration sample delivery system to determine a suite of environmentally-significant elements in potable and waste waters according to US EPA Method 200.7.

EPA Method 200.7 (2) covers the use of radial or axial ICP-OES for the determination of 23 metals and some non-metals in potable waters and wastewaters for regulatory compliance. This method contains a lengthy description of procedures for the collection, preservation, and preparation of samples for analysis. The objective of this study was to follow the method using the PerkinElmer® Optima 8300 ICP-OES coupled with the prepFAST™ automated auto-dilution/calibration system (ESI, Omaha, Nebraska). The prepFAST system provides a number of advantages over conventional ICP-OES introduction systems, the most significant being higher sample throughput and reduced memory effects. The system allows accurate, syringe-driven auto dilution of samples and standards, eliminating manual dilution errors and increasing calibration range. By reducing consumption of the sample and high-purity chemical reagents, the prepFAST™ Auto-Dilution/Calibration System is one of the most cost-effective ways of improving laboratory productivity (3).

Experimental

To be compliant with EPA Method 200.7, a number of quality checks must be carried out before sample analysis can be started. Some of these tests include instrument tuning; calibration; performance checks, including interelement correction (IEC); instrument detection limits (IDL); method detection limits (MDL); quality control (QC) samples; linear dynamic range (LDR); and method blanks and spike recoveries. The instrument was set up with the operating conditions shown in Table I using the recommended wavelengths and background correction points described in the method. For some analytes, multiple wavelengths were used, based on sensitivity and/or freedom from interference. It should also be emphasized that a plasma flow rate of 8 L/min was used for this analysis, which is typically 50% lower than other ICP-OES systems.

Table I

Calibration

The majority of solutions prepared for this study were diluted in 2% nitric acid. For calibration, a 1 mg/L and a 5 mg/L standard were prepared and the prepFAST system was used to automatically dilute these solutions to prepare other calibration standards and blanks. The prepFAST system was also set up to automatically dilute samples, some of which were deliberately prepared over the range of the calibration curve, so it would automatically carry out a sample dilution. Yttrium was used as the internal standard for the analysis of all samples.

Results

It is not possible to show the complete set of sample and QA/QC data in this application brief. However, to emphasize the accuracy of the dilutions carried out by the prepFAST, Table II shows sample analysis data for the recovery of the initial calibration verification (ICV) solution (referenced to 10x user-prepared dilution) with further prepFAST dilutions at 10x or 100x, when necessary. It can be clearly seen that the recoveries of all analytes are well within the EPA guidelines of 85–115%. For the complete set of sample results generated in this study, please refer to the published application note (1).

Table II

Conclusion

This study has demonstrated the capabilities of the PerkinElmer Optima 8300 ICP-OES coupled with the prepFAST™ system to produce results that meet the rigorous requirements outlined in US EPA Method 200.7 for the analysis of potable and waste waters. The accuracy and precision of the instrument allows less time to be spent on meeting the performance requirements. The prepFAST™ system simplifies sample preparation, allowing higher sample throughput, reducing memory effects, and minimizing errors and contamination.

References

(1) D.K. Bradshaw and L. Thompson, "The Analysis of Water and Wastes by U.S. EPA Method 200.7 Using the Optima 8300 ICP-OES and prepFAST Auto-Dilution/ Calibration System," PerkinElmer Inc. Application Note, 2013.

(2) EPA Method 200.7 Revision 4.4, "Determination of Metals and Trace Elements in Water and Wastes by ICP-AES." http://water.epa.gov/scitech/methods/cwa/bioindicators/upload/2007_07_10_methods_method_200_7.pdf.

(3) "PrepFAST On-line Auto-dilution and Auto-calibration Sample Delivery System: Elemental Scientific Inc." Application Note, http://www.icpms.com/products/prepfast.php.

PerkinElmer, Inc.

940 Winter Street, Waltham, MA 02451

tel. (800) 762-4000 or (203) 925-4602

Website: www.perkinelmer.com


Related to this article

Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.
Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.