Wire Grid Polarizers for IR Spectroscopy

Broadband wire grid polarizer (WGP) products suitable for MWIR and LWIR (mid- and long-wavelength infrared) spectroscopic applications requiring high contrast were developed on thin, AR-coated silicon substrates using Moxtek wafer-scale aluminum Nanowire? patterning capabilities. The Moxtek WGP performance was characterized from the SWIR to LWIR and shows marked improvement over competing WGP products, especially in terms of the contrast ratio between passing and blocking state polarizer transmittance.

Broadband wire grid polarizer (WGP) products suitable for MWIR and LWIR (mid- and long-wavelength infrared) spectroscopic applications requiring high contrast were developed on thin, AR-coated silicon substrates using Moxtek wafer-scale aluminum Nanowire® patterning capabilities. The Moxtek WGP performance was characterized from the SWIR to LWIR and shows marked improvement over competing WGP products, especially in terms of the contrast ratio between passing and blocking state polarizer transmittance.

Polarizers are used in IR spectroscopy for characterization of dielectric and metallic coatings, analysis of birefringent materials, and for indexing the vibrational modes in crystals and textured films. A real polarizer has a finite contrast ratio, thus some light with the unwanted polarization state is transmitted, which can impact measurement accuracy. The wire grid polarizer (WGP) is often preferred for IR applications, and consists of an array of metallic lines with sub-wavelength pitch, often supported by a transparent substrate. Benefits of the WGP over competing designs include a compact form factor with improved passing state transmittance, minimal performance variation with angle of incidence or wavelength, and improved stability in high temperature and high brightness environments. Unfortunately, WGP products designed for MWIR and LWIR applications have historically suffered from low contrast between transmission of linearly polarized light oriented in the passing and blocking states, which is mostly due to their relatively large wire grid pitch (typically ≥ 370 nm). By dramatically reducing the pitch from that found in typical IR WGP products to 144 nm, Moxtek has developed MWIR and LWIR polarizers with greatly improved contrast.

Experimental Conditions

Transmission experiments were performed using both Nexus 870 and CARY 670 FTIR spectrometers. Rotation of the WGP sample with respect to a fixed pre-analyzer allowed for transmittance measurements in the passing and blocking states. The pre-analyzers consisted of two aligned Moxtek WGP's (both either MWIR or LWIR) separated by a gap of more than 100 μm. Open and blocked beam configurations and a silicon reference standard were used to validate instrument accuracy for transmission measurements.

Results

Moxtek's MWIR polarizer typically transmits better than 95% of the passing state between 3.5 and 5.5 μm while maintaining a contrast ratio of better than 35 dB from 3.0–5.5 μm. The Moxtek LWIR polarizer typically transmits better than 68% of the passing polarization state between 7 and 15 μm and has a contrast ratio exceeding 38.5 dB. Table I compares the performance of the 144 nm pitch Moxtek products to a 250 nm pitch WGP from another supplier and clearly demonstrates the dramatic improvement in contrast ratio.

Table I: Performance comparison between Moxtek infrared wire grid polarizers on AR-coated silicon and a competing product on thallium bromoiodide (KRS-5) substrate

Conclusions

Wire grid polarizers have many advantages over other designs when used as analyzers in infrared spectroscopic applications, but have historically suffered from poor contrast between transmittance in the passing and blocking states. To address these limitations Moxtek has extended its aluminum Nanowire® patterning capabilities to AR-coated silicon substrates designed for the mid- and long-wavelength infrared. These new product offerings offer extraordinary improvements in contrast while maintaining excellent efficiency in passing state transmittance.

Moxtek

452 West 1260 North, Orem, UT 84057

tel. (800) 758-3110

Email: info@moxtek.com

Website: www.moxtek.com


Related to this article

Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.