Determination of Mercury in Water Samples Using Cold Vapor Atomic Absorption Spectroscopy — EPA 245.1

With over 1300 commercial testing labs in the United States, one out of every three samples is for water analysis. There has been an even stronger emphasis on mercury testing and its concentration in our water supplies.

With over 1300 commercial testing labs in the United States, one out of every three samples is for water analysis. There has been an even stronger emphasis on mercury testing and its concentration in our water supplies. This application note describes the use of cold vapor atomic absorption spectroscopy as described in the EPA method 245.1 to determine mercury concentration in water samples including drinking, surface, ground, sea, brackish waters, and industrial and domestic waste waters.

Experimental

Determine mercury concentrations in surface water samples. In addition to the water samples, pre and post QC checks (15 ppt standards) served to ensure instrument stability and accurate results.

Principle of Operation

Milestone's Mercurio AAMA-80 incorporates the following sequence to perform a direct measurement in a liquid sample: Aliquot of sample transferred automatically to reaction mixing block through a peristaltic pump where it is mixed with argon gas and a stannous chloride reagent solution, mercury vapors pass through a gas-liquid separator and purification system followed by Hg amalgamation and detection. Following amalgamation, the mercury vapors are released to a single beam, fixed wavelength atomic absorption spectrophotometer. Absorbance is measured at 253.7 nm as a function of mercury content. The system is concurrently cleaned with the amalgamation step for a seamless and more productive process.

Figure 1: Milestone’s Mercurio AAMA-80.

Results

See Table I.

Table I: QC Hg standards and surface water samples

Discussion and Conclusion

The Mercurio AAMA-80 successfully recovered 100% of the 15 ppt Hg standards and processed 10 surface water samples with excellent reproducibility and accuracy without any interference.

Milestone Inc.

25 Controls Drive, Shelton, CT 06484

tel. (866) 995-5100, fax (203) 925-4241

Website: www.milestonesci.com

Articles in this issue

almost 13 years ago

PIKE Technologies

almost 13 years ago

Amptek, Inc.

almost 13 years ago

Ocean Optics

almost 13 years ago

Photonis USA

almost 13 years ago

Shimadzu Scientific Instruments

almost 13 years ago

Milestone, Inc.

almost 13 years ago

Savillex Corporation

almost 13 years ago

WITec GmbH

almost 13 years ago

Thermo Fisher Scientific

Related to this article

Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.
Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.