A Diamond ATR Study of Worn Surfaces on Metal

Tribology–the study of friction, wear, and lubrication–is important in mechanical, electromechanical, and biological systems. Changes in materials due to abrasion, friction, lubrication, erosion, and corrosion result in failures of the moving components during use. These effects are usually minimized by careful surface finish design and lubrication. Recycling and refinishing surfaces can extend the lifetimes of components. By examining worn surfaces, better methods can be developed to prevent wear. This applications note uses infrared spectroscopy to study tribology changes in steel.

Experimental

The sample investigated was 28 mm in diameter by 8 mm thick with a carbon-based coating. The sample was lubricated, tested, and then cleaned with n-heptane, leaving chemisorbed species in the wear scratches on the surface.

The spectra of the disc were measured on an FT-IR spectrometer equipped with the Harrick VideoMVP™ (see Figure 1). The background spectrum was collected from the clean ATR crystal. Then the sample was placed, face down, on the diamond crystal and positioned over the area of interest using the image on the display. The pressure applicator was lowered to apply ~240 N of force to compress the sample against the crystal. The spectrum and image were then recorded. All spectra were collected from 4000 to 400 cm-1 using 32 scans at 8 cm-1 resolution.

Results and Discussion

Figure 2 shows spectra measured from a section with wear scratches in comparison to a shiny smooth section, along with photographs of the areas of each sample that were examined. The infrared spectrum of the wear scratches show a weak broad band in the –OH stretching region and additional bands at 1450 cm-1, 1165 cm-1, 943 cm-1, and 605 cm-1, all of which are not present in the unworn section. This indicates that a residue from the lubricant or chemisorbed species remains in the worn area. More detailed analysis could be used to identify the chemisorbed species.

Conclusion

The results presented here clearly show that diamond ATR can be used to examine tribology materials. The viewing capabilities and small sampling area of the VideoMVP allow for inspection of the wear scratches in comparison to the unworn areas.

Harrick Scientific Products, Inc.
141 Tompkins Ave., Pleasantville, NY 10571
tel. (914) 747-7202, fax (914) 747-7409
Website: www.harricksci.com


Related to this article

Pittsburgh, Pennsylvania, USA. Beautiful Pittsburgh at twilight. | Image Credit: © checubus - stock.adobe.com
Organized by CPACT, the April 27 to 29 meeting at the Marriott Pittsburgh City Center will bring together scientists, engineers, and vendors across the process industries, with keynotes on PAT, Raman, and NIR applications and an abstract deadline of December 7, 2026.
Abstract concept of connecting molecules DNA network vector neurons | Image Credit: © bill2499 - stock.adobe.com
Garth Simpson of Purdue University outlines plans to survey more ordered molecular assemblies and to apply interface-specific chiral spectroscopy to enzyme reactors, solid-phase synthesis, spin-selective reactions, and, ultimately, a new analytical method.
Curated 2026 Evergreen Content from Spectroscopy. ©  Chanakarn -chronicles-stock.adobe.com
Spectroscopy's editors gather 143 evergreen articles from 2026, including tutorials, terminology guides, technique reviews, application surveys, peer-reviewed research and the magazine's long-running columns, and organize them by subject so readers can build lasting knowledge one topic at a time.