Digestion of Difficult APIs in Accordance with USP

The new USP Chapters <232> and <233> will become official on January 1, 2018 and bring significant changes in sample preparation and analysis of pharmaceutical samples both in excipients and APIs. Certain APIs pose challenges to acid digestion because they are very stable compounds and not easily broken down.

USP methods <232> and <233> call for total digestion of pharmaceutical samples and quantification of individual elements. Many pharmaceutical materials can be easily digested but APIs with multiple aromatic ring structures can be very difficult to completely break down and obtain a clear digestion, as prescribed in the new chapters. Microwave digestion can provide a solution to many of these compounds.

This application note will focus on the use of the CEM MARS 6 microwave digestion system and new high performance iPrep vessels to completely digest very difficult APIs. The ring structures of the APIs are shown to illustrate the high stability of these compounds. The sample sizes listed represent the maximum allowable in order to achieve a complete digest.

Experimental Conditions

Three APIs were prepared using a CEM MARS 6 microwave digestion system equipped with iWave technology and iPrep vessels.

iWave is a novel Light Emitting Technology(tm) that measures the temperature of the actual sample solution inside the vessel and does not require an internal probe.

iPrep is a simple to use three-piece vessel which uses a simple hand torque device to seal. The patented dual-seal design provides for higher temperatures and fine control of the vent and reseal process necessary for these sample types.

Each API sample was weighed and added to an iPrep liner with 9 mL HNO3 and 1 mL HCl. The vessels were capped, assembled, and placed in the MARS 6 for digestion. Digestion parameters are recorded in Table I.

Results

The MARS 6 with iPrep vessels was able to completely digest each of the API materials and provide a clear solution. Of note is that the TrixiPhos digestate is a bright yellow color. This is a result of the Pd catalyst used when synthesizing the compound. Each API sample was run in duplicate in order to confirm the success of the sample preparation. All samples were completely digested as shown in Figure 1.

Conclusions

The MARS 6 system with iPrep vessels is an ideal option for working with these difficult pharmaceutical materials.

CEM Corporation
P.O. Box 200, Matthews, NC 28106
tel. (704) 821-7015, (800) 726-3331
Website: www.cem.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.