Performance Characteristics of a PFA Cyclonic Spray Chamber with HR ICP-MS

Elemental ratio determinations in foraminiferal calcite require the measurement of trace levels of Na, Mg, Sr, and ultra-trace levels of Li, B, Ba, Cd, Mn, U in a calcium rich matrix and ratioed to calcium.

The Savillex PFA cyclonic spray chamber was used with a Thermo ElementXR HR-ICP-MS for the measurement of trace element/calcium ratios in foraminiferal calcite. The signal stability and washout (particularly for boron which is a "sticky" element) were studied. The spray chamber was found to have excellent sensitivity, good short and long term stability, and fast washout (60 s for B), making it well suited to the application.

Elemental ratio determinations in foraminiferal calcite require the measurement of trace levels of Na, Mg, Sr, and ultra-trace levels of Li, B, Ba, Cd, Mn, U in a calcium rich matrix and ratioed to calcium. Samples are run in an 0.1 M HNO3/0.3 M HF matrix to prevent boron carryover between samples, so an inert sample introduction system is required. PTFE and PFA inert spray chambers generally have inferior performance to quartz spray chambers. We evaluated the recently introduced Savillex PFA cyclonic spray chamber to see if advancements in technology have improved the performance of inert sample introduction systems for ICP-MS. A Thermo ElementXR HR-ICP-MS was fitted with a Savillex inert kit comprising of an inert torch with platinum injector, PFA cyclonic spray chamber (no surface treatment) with baffle, and a C100 PFA concentric nebulizer.


Figure 1: Savillex PFA cyclonic spray chamber.

An acid blank and three standards (standards 0, 4, and 8) containing 10 ppm Ca and varying ppt–ppb levels of trace elements were run. After the initial data acquisition for all standards, the blank, standard 0, and standard 8 were measured 17 times each in sequence, to check for reproducibility and drift. The spray chamber was found to have good short-term stability with RSDs of 0.5–1%, and good long-term reproducibility RSDs of 1.5–2.9% over 4 h. No drift correction was applied. Washout of B from the high standard (0.74 ppb) to blank levels took <60 s, demonstrating the good washout characteristics of the spray chamber.


Figure 2: Boron signal over 5 h-acid blank and standard 0 (2 ppt B in 10 ppm Ca), showing excellent sensitivity and washout (the higher standards in between have been excluded for clarity). For full details visit www.savillex.com/ICP.

Savillex Corporation
10321 West 70th Street, Eden Prairie, MN 55344-3446
tel. 952-935-4100
Website: www.savillex.com


Related to this article

Scientist With Portable Spectrometer in Natural Field Setting ©  By Tika -chronicles-stock.adobe.com
The bulky bench-top NIR spectrometer is quietly being dismantled and rebuilt as a wafer-scale photonic chip, a self-calibrating algorithm, and a sensor small enough to ride in a shirt pocket. What once demanded a grating, a moving mirror, and a climate-controlled lab now fits inside a handheld module, a bioreactor probe, or a drone payload, and it increasingly figures out what it is looking at on its own.
Sizing Up the Nanoscale: Measuring Nanocluster Aerosol in Indoor Air
In the first part of a multi-part Q&A, Brandon Boor, the Dr. Margery E. Hoffman Associate Professor in the Lyles School of Civil and Construction Engineering at Purdue University, describes the instrumentation and methodology behind measuring nanoparticle size distributions at the nanocluster scale (1–3 nm) and outlines the technical challenges of acquiring reliable, real-time data at these dimensions.
Human body wireframe on glowing platform undergoing futuristic body scan. © sergray(noAIelemens) -chronicles-stock.adobe.com
Jurgen Popp, Thomas Mayerhofer, and colleagues at Leibniz IPHT and Friedrich Schiller University Jena introduce the Personalized Optical Digital Twin (PODT), a Photonics21 contribution to Europe's Virtual Human Twin ecosystem that connects molecular photonics—Raman blood analysis, coherent Raman tissue imaging, and multimodal endomicroscopy—with longitudinal physiology and clinical data. Drawing on the published multicenter INTELLIGENCE trials, the authors argue that technical feasibility and clinical utility must be evaluated separately as the field moves toward Europe's FP10 research agenda.