Trace-Level Pesticide Detection Utilizing Surface-Enhanced Raman Spectroscopy (SERS)

Author(s)Ocean Insight

Several pesticides were detected at parts-per-million levels on apple skin using a unique swabbing method and surface-enhanced Raman spectroscopy (SERS).

Commonly used pesticides, such as organothiophosphates and fungicides, can attack the central nervous system, posing a risk to humans and other animals upon exposure (1). SERS has been investigated as a technique to detect trace levels of pesticides on food items, and offers reduced instrument size, fast measurement times, non-destructive sampling, and simple implementation compared to the traditionally used methods, such as HPLC and GC–MS (2). In this study, we utilize a swabbing technique that has the potential to be a testing platform for field use to detect trace levels of pesticides on apple skin.

Experimental Conditions

Pesticide-containing apple skin pieces are swabbed and added to colloidal gold nanoparticles. The mixtures are interrogated with 785 nm excitation laser with 3 s integration time and 350 mW laser power, and the Ocean Insight QE Pro spectrometer.

Results

The pesticides examined in this study were thiram, malathion, acetamiprid, and phosmet. The SERS spectra of each pesticide after being applied to apple skin and swabbed are presented in Figure 1. The observed peaks are consistent with reports in the literature.


Figure 1: SERS spectra of swabs of 1 ppm (a) malathion, (b) acetamiprid, (c) phosmet, and (d) thiram on apple skin.

Conclusions

We present a swabbing technique that has the potential to be a testing platform for field use and utilizes colloidal gold to detect trace levels of several pesticides on apple skin. This technique can detect each pesticide down to 1 ppm, where the pesticide residue tolerances on apples are 5 ppm, 8 ppm, 1 ppm, and 10 ppm for thiram, malathion, acetamiprid, and phosmet, respectively (3). The results presented here indicate that SERS coupled with the swab method is a valuable tool, and has significant potential for identifying pesticide residues on the surface of fruits for food quality and safety control.

References

(1) M. Stoytcheva, InTech, 30–48 (2011). doi: 10.5772/1003.

(2) M.L. Xu, Y. Gao, X.X. Han, and B. Zhao, J. Agric. Food Chem. 65, 6719–6726 (2017).

(3) Office of the Federal Register, Electronic Code of Federal Regulations, 2018, https://www.ecfr.gov (8 January 2019).

Ocean Insight
8060 Bryan Dairy Road, Largo, FL 33777
tel. +1 (727) 733-2447
Website: www.oceaninsight.com


Related to this article

Spectroscopy Top 10 DOI Articles of the Month (October 2026)
The following articles are the most accessed digital object identifier (DOI) manuscripts for Spectroscopy and LCGC International during the month of October 2026. Nine articles are individually ranked here by DOI page views; the seventh-highest entry (99 views) was recorded against the bare LCGC North America journal-level DOI rather than an individual article and is therefore not attributable to a single manuscript.
Mapping Molecular Mobility Pixel by Pixel
Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.
Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.