Raman spectroscopy is an excellent technique for the identification and characterization of fuels. With no requirement for sample preparation and the power to identify and quantify materials, Raman has many uses across a range of industries.
Raman spectroscopy is an excellent technique for the identification and characterization of fuels. With no requirement for sample preparation and the power to identify and quantify materials, Raman has many uses across a range of industries.
Biodiesel is a nonpetroleum-based diesel fuel made from vegetable oil or animal fats. Biodiesel can be used in most diesel engines with little to no engine modification required, and can be blended with petroleum diesel fuel to provide a cleaner burning, lower emission diesel fuel.
As biodiesel production increases, Raman spectroscopy will become a useful analytical tool during the refining process to assess incoming raw materials, monitor the production process, and confirm the quality of the final product.
Today's Raman measurement options range from handheld and benchtop systems such as the Ocean Optics IDRaman series, to modular, "build your own" Raman systems including spectrometers, lasers, probes, and sample holders.
The Raman measurements described here were made with a modular setup comprising the Maya2000 Pro-NIR spectrometer, a 785 nm Raman laser, a Raman-coupled probe for 785 nm Raman and a sample holder. Acquisition parameters were a 500 ms integration time with no scans averaged and no boxcar smoothing. Samples of corn oil (sometimes used as a diesel alternative) and petroleum-based diesel were placed in small glass vials for analysis.
The Raman spectra for corn oil and diesel are shown in Figure 1. While these spectra share some features due to the hydrocarbon content of the samples, there are several spectral differences observed in the fingerprint region from 500–2000 cm-1. Even though both samples are suitable for use as fuel in diesel engines, they have distinct Raman spectra that distinguish the corn oil-based biodiesel fuel from the petroleum-based diesel fuel. Note that in addition to identifying the fuel type based on its Raman fingerprint, one could obtain more quantitative information from the spectra by applying the appropriate chemometrics models.
Figure 1: Raman is an excellent method for distinguishing corn oil-based biodiesel fuels from petroleum-based diesel fuels.
One notable difference for the spectrum of corn oil is the presence of stearate (a form of fatty acid found in animal and vegetable fats and oils) in the region from 1600–1800 cm-1. While stearate artifacts and other spectral differences allow for easy discrimination of these fuels, samples with even more closely aligned spectral peaks can be distinguished. Indeed, by using a narrower slit in the spectrometer optical bench, you will achieve a higher Raman shift resolution over a narrower spectral range.
With their unique hydrocarbon composition, fuels are well suited for identification and characterization using Raman analysis. The wealth of spectral features in the Raman spectra for fuels can be used in a range of applications including determination of critical fuel parameters, fuel classification, and detection of counterfeit fuels.
With all the choices available, the modular approach to Raman measurements provides a nearly endless choice of setups with the ability to change or add components to meet your evolving measurement needs.
Ocean Optics
830 Douglas Avenue, Dunedin, FL 34698
tel. (727) 733-2447, fax (727) 733-3962
Website: www.OceanOptics.com
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
The World of Microplastics Up to Date – an Overview
April 23rd 2024Watch this 20-minute educational video by Andreas Kerstan, Agilent Product Specialist in molecular spectroscopy, to gain a comprehensive update on the microplastics landscape and the environmental concerns related to them. Discover the current challenges in microplastics characterization and how Agilent innovative solutions and techniques, including FTIR, LDIR, GC/MS, and ICP-MS, are addressing these issues head-on.