Diamond ATR-FTIR Study of Nitriles
Experimental
Infrared spectra were collected on an FT-IR spectrometer equipped with the Harrick DiaMaxATR™ single-reflection high-throughput diamond ATR accessory or ConcentratIR2 multiple-reflection diamond ATR (see Figure 1). The system was purged to remove water vapor and CO2. Spectra were collected at 8 cm-1 resolution, and signal averaged over 32 scans. The spectra were referenced to the clean ATR crystal.
The sample used here was a High Five Nitrile Disposable Glove (lightly powdered, textured, P/N N842). For the single-reflection ATR measurement, a portion of the glove was pressed against the ATR crystal using the maximum force supplied by the build-in pressure applicator. For the multiple-reflection ATR measurements, the sample was gently pressed against the ATR crystal with compressible foam to fill the trough, and to apply force as uniformly as possible.
Results and Discussion
Figure 2 shows the two spectra. The upper spectrum measured with the multiple-reflection ATR shows only noise in the 2300–1900 cm-1 region, due to the high absorbance of diamond in that region. The lower spectrum, however, clearly shows the C≡N stretch at 2238 cm-1. Another weaker band is also apparent in the single-reflection ATR spectrum at 2162 cm-1, possibly indicating the presence of an alkyne group.
Conclusion
The Harrick DiaMaxATR high-throughput, single-reflection diamond ATR is effective at measuring functional groups that are frequently obscured by the strong diamond lattice bands in the 2300–1900 cm-1 region. This makes it more suitable for analysis of the function groups in that region, including nitriles, isocyanates, isothiocyanates, diimides, azides, and ketenes.
Harrick Scientific Products, Inc.
141 Tompkins Ave., Box 277, Pleasantville, NY 10570
tel. (914) 747-7202, fax (914) 747-7409
Website:
Articles in this issue
almost 6 years ago
Non-Destructive FT-IR Measurements via Diffuse Reflection Samplingalmost 6 years ago
Tri-Range Applications of the Spectrum 3 Infrared Spectrometeralmost 6 years ago
Raman Spectroscopy as a Tool for Studying Polymer Phase Transitionsalmost 6 years ago
Raman Imaging for the Analysis of Food Productsalmost 6 years ago
A Sampling Flexibility for Raman Spectroscopyalmost 6 years ago
Identifying Textiles with Extended-Range NIR Spectroscopyalmost 6 years ago
Analysis of Gemstones Using EDXRFRelated to this article