Microwave Digestion of Pharmaceutical Samples for USP Method

Pharmaceutical samples spiked with several heavy metals were prepared for ICP-MS analysis using microwave digestion following the protocols in proposed USP chapters <232> and <233>. The results of the spike recovery study are discussed.

Pharmaceutical samples spiked with several heavy metals were prepared for ICP-MS analysis using microwave digestion following the protocols in proposed USP chapters <232> and <233>. The results of the spike recovery study are discussed.

The current United States Pharmacopeia (USP) chapter <231> Heavy Metals will be replaced in 2014 by USP chapter <232> Elemental Impurities – Limits and USP chapter <233> Elemental Impurities – Procedures. These new methods will provide a more accurate analysis by identifying specific elements of interest and will enable the detection of additional elements, such as chromium and catalyst metals, which may be introduced during the manufacturing process. Additionally, the new methods require closed vessel microwave digestion for samples that cannot be successfully dissolved in an aqueous solution or an organic solvent, preventing the loss of volatile analytes from the open flask heating technique used in Method <231>. The digested materials can then be analyzed by ICP-OES or ICP-MS, depending upon the detection limits required.

CEM's MARS 6 and Discover SP-D microwave sample preparation systems were used to prepare several finished pharmaceutical products. The MARS system has the ability to prepare up to 40 samples simultaneously, while the Discover SP-D with an autosampler can prepare a sample every 10 min completely unattended. The accuracy and precision of the spiked recovery results for each system are discussed.

Experimental

In our study, pharmaceutical finished products including a tablet with the active ingredient acetylsalicylic acid, a tablet with the active ingredient loratadine, and liquid gel capsule with the active ingredient diphenhydramine HCl were digested and analyzed with an ICP-MS. A tablet or liquid gel capsule was added to a MARS EasyPrep Plus vessel or a Discover SP-D 35-mL vessel along with 9 mL of concentrated nitric acid and 1 mL of spiked DI water containing 50 ppb of arsenic, mercury, and cadmium. The liquid gel capsule also required hydrogen peroxide in order to fully oxidize the sample – 1 mL in the EasyPrep Plus vessel and 0.5 mL in the 35-mL vessel. Each of the samples and blanks was prepared in triplicate using the digestion parameters shown in Table I.

Table I: Digestion parameters for MARS 6 and Discover SP-D

The digested samples were then diluted with DI water to a volume of 50 mL and analyzed with an ICP-MS.

Results

The accuracy of proposed USP Method <233> is 70–150% with a relative standard deviation of not more than 20%. The results of the digestions performed with the MARS 6 and Discover SP-D are shown in Table II. The volatile elements, arsenic and mercury, achieved percent recoveries of 101–136% and an RSD of less than 15.1%. The non-volatile element, cadmium, showed percent recoveries from 94–111% and an RSD of less than 9.1%. The percent recovery and RSD for each element is within the proposed requirements of USP Method <233>.

Table II: Recovery results of 50 ppb spike of As, Hg, Cd in finished pharmaceutical products

Conclusions

The CEM MARS 6 and the Discover SP-D provide a fast, accurate means to prepare pharmaceutical ingredients and finished products to meet the requirements of the upcoming USP Method <233>.

CEM Corporation

P.O. Box 200, Matthews, NC 28106

tel. (800) 726-3331

Website: www.cem.com/usp


Related to this article

Optical and Electronic Instrument Components,  ©  ridwan -chronicles-stock.adobe.com
The updated and expanded 2026 edition of Spectroscopy's Instrument Components Terminology Guide defines the sources, detectors, optical and dispersive elements, sampling hardware, physical constants, and unit conversions used across UV-vis, NIR, IR, Raman, X-ray, and atomic spectroscopic instrumentation, with new terms reflecting component developments reported between 2022 and 2026.
Pittsburgh, Pennsylvania, USA. Beautiful Pittsburgh at twilight. | Image Credit: © checubus - stock.adobe.com
Organized by CPACT, the April 27 to 29 meeting at the Marriott Pittsburgh City Center will bring together scientists, engineers, and vendors across the process industries, with keynotes on PAT, Raman, and NIR applications and an abstract deadline of December 7, 2026.