|Articles|February 1, 2021

Application Notebook

  • The Application Notebook 02-01-2021
  • Volume 36
  • Issue s2

See-Through Measurements of Illicit Substances in Commercial Containers with the TacticID®-1064 ST

Experiment
The TacticID-1064 ST is equipped with a see-through (ST) mode scan function that allows users to identify chemicals behind thick and opaque barriers with the use of an ST sampling adapter. A hit quality index (HQI) is used to match the unknown sample to a library spectrum. The HQI calculation ranges from 100 (best match) to 0 (worst match). The system employs an automatic integration time. The laser power is adjustable, but was set to 90% for these measurements. The number of hits can also be adjusted.

Materials tested:

  • Fentanyl: highly toxic opioid that is often mixed with heroin and other street drugs
  • N-acetylanthranilic acid: Drug Enforcement Agency List I controlled drug precursor, used in synthesis of methaqualone, highly fluorescent with 785 nm excitation
  • Caffeine: stimulant often used as a cutting agent in drug manufacturing

Containers:

  • manila envelopes
  • padded shipping packages
  • high-density white polyethylene (HDPE) bottles

Results

Fentanyl citrate powder inside a thin plastic bag was placed inside a manila envelope and tested with the TacticID-1064 ST. Fentanyl citrate was successfully identified directly through the manila envelope with an HQI of 85.0. N-acetylanthranilic acid is a light brown compound that is used in the synthesis of methaqualone and mecloqualone, both Schedule I drugs. According to the International Narcotics Control Board (INCB), 10.4 metric tons of N-acetylanthranilic acid have been seized globally since 2000. When measured with a 785 nm laser, the Raman signal is completely overwhelmed by the generated fluorescence (Figure 1, red trace), making identification impossible with Raman. The 1064 nm laser of the TacticID-1064 ST does not generate fluorescence (Figure 1, blue trace), and a good Raman spectrum can be collected and used for identification against the spectral library.

The ST mode on the TacticID-1064 ST was used to measure N-acetylanthranilic acid through a white plastic (HDPE) bottle with an HQI of 92.2.


Conclusion

The TacticID-1064 ST puts operator safety first, removing the need to actively sample from opaque packaging to identify illicit substances. The 1064 nm laser excitation removes fluorescence issues generally associated with 785 nm
Raman systems.

B&W Tek, LLC
19 Shea Way, Newark, DE 19713
Tel: +1 (302) 368-7824, Fax: +1 (302) 368-7830
Website: www.bwtek.com


Related to this article

Education distance learning, e-learning, and online learning are all types of personal development. | Image Credit: © ipuwadol - stock.adobe.com
In this segment of our SciX interview with Kevin Wilkinson, a professor of environmental chemistry at the Université de Montréal, discusses extending lead isotope work to greater distances and explains why graduate students need adaptability, critical judgment, and coding skills more than expertise in any single instrument.
Atoms Don't Lie–The Remaking of Atomic Spectroscopy, © Hallie -chronicles-stock.adobe.com
Atomic spectroscopy has quietly gone from counting elements in bulk to counting single nanoparticles, imaging tissue at the micrometer scale, and reading Martian rocks from several meters away. The instruments got smaller, the software got smarter, and the questions they can now answer got a lot more interesting.
Tutorial Tuesday: When to Use Infrared Spectroscopy
In this “Tutorial Tuesday” clip, Peter Larkin, who is a Manager of Spectroscopy, Thermal Analysis, and Chromatography at Syensqo, highlights how people use the method or technique that they are comfortable with, even if it’s not the best one to use.