Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions, both aqueous and non-aqueous. Most multiple ATR accessories use ATR crystals which are susceptible to wear and damage. Diamond, although an excellent choice due to its refractive index, high degree of chemical inertness, and low coefficient of friction, is generally not used because of its expense and its strong lattice bands around 2200 cm-1.
S.L. Berets1, M. Nassar1, and M. Milosevic2, 1Harrick Scientific Products, Inc. and 2MEV Photonics
Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions, both aqueous and non-aqueous. Most multiple ATR accessories use ATR crystals which are susceptible to wear and damage. Diamond, although an excellent choice due to its refractive index, high degree of chemical inertness, and low coefficient of friction, is generally not used because of its expense and its strong lattice bands around 2200 cm-1. However, by using a small ATR crystal and a short pathlength through the crystal, as is commonly done with single reflection diamond ATR accessories, these effects can be minimized. This note explores the use of a multiple reflection diamond ATR accessory to distinguish between several different aqueous liquids by examining low concentration functional groups.
Harrick Diamond ConcentratIR™ (see Figure 1), which supplies seven reflections from the sample, was installed and aligned in a commercial FT-IR spectrometer. Spectra were collected at an 8 cm-1 resolution and signal averaged over 64 scans. Three different eye drops were examined: Visine®, Sun Mark®, and Patanol®. A single drop of each was placed on the ATR crystal for analysis.
Figure 1: ConcentratIR⢠multiple reflection ATR.
Figure 2 shows the region with the greatest spectral differences and many of these variations can be directly attributed to the published components. For example, the peak at 1400 cm-1 which appears in Visine represents C–O–H in-plane bending, characteristic of an acid functional group. Visine is the only formulation that contains boric acid, thus it is consistent with the presence of this band only for this substance. The same is true of the peak at 1150 cm-1 in the Visine spectrum, which denotes C–O bond carboxylic acid stretch. The Patanol spectrum shows only distinct peak, which occurs around 1075 cm-1, and it is likely to be the N–C stretching of its main component: olopatadine. The SunMark spectrum contains a C–O stretching peak at 1030 cm-1 characteristic of an alcohol, as expected from glycerin. It also exhibits a peak near 1110 cm-1 due to N–C stretching in its only active ingredient ketotifen.
Figure 2: Spectra of all samples in the 1500 cm-1 to 1000 cm-1 region. Patanol® (red), SunMark® (black), and Visine® (blue).
The ConcentratIR™, a multiple reflection diamond ATR accessory is effective in revealing spectral differences between complex solutions, even if those solutions are over 95% water. Its seven reflections in a 4 mm in diameter sampling area make it ideal for analyzing small quantities of liquids.
Harrick Scientific Products, Inc.
141 Tompkinss Ave., Box 277, Pleasantville, NY 10570
tel. (914) 747-7202, Fax: (914) 747-7209
Website: www.harricksci.com
Single Cell and Microplastic Analysis by ICP-MS with Automated Micro-Flow Sample Introduction
April 25th 2024Single cell ICP-MS (scICP-MS) is increasingly seen as a powerful and fast tool for the measurement of elements in individual cells, mainly due to the high sensitivity and selectivity of ICP-MS. Analysis is performed in the same way as single nanoparticle (spICP-MS) analysis, which has become a well-established technique for the analysis of nanoparticles and particles.
Hot News on Agilent LDIR, New Developments, and Future Perspective
April 25th 2024Watch this video featuring Darren Robey and Dr. Wesam Alwan from Agilent Technologies to gain insights into the future trends shaping microplastics research and the challenges of their characterization. Discover the essential components necessary for accurate microplastics analysis and learn how the Agilent 8700 LDIR system addresses these challenges. Offering rapid and precise analysis capabilities, along with easy sample preparation methods that minimize contamination, the Agilent 8700 LDIR system is at the forefront of advancing microplastics research.
The World of Microplastics Up to Date – an Overview
April 23rd 2024Watch this 20-minute educational video by Andreas Kerstan, Agilent Product Specialist in molecular spectroscopy, to gain a comprehensive update on the microplastics landscape and the environmental concerns related to them. Discover the current challenges in microplastics characterization and how Agilent innovative solutions and techniques, including FTIR, LDIR, GC/MS, and ICP-MS, are addressing these issues head-on.