An Infrared Investigation of Nail Polish Remover Using Multiple Reflection Diamond ATR

Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions. If one component is volatile, trace analysis can be simplified by allowing the volatile liquids to evaporate.

Multiple reflection ATR is ideal for detecting miniscule levels of components in solutions. If one component is volatile, trace analysis can be simplified by allowing the volatile liquids to evaporate. A multiple reflection ATR with a small sampling area is advantageous in cases where limited sample is available, as it allows additional concentration of the residue.

This note explores the use of a multiple reflection diamond ATR accessory to detect and identify the minor chemical variability between different types of nail polish remover.

Experimental

Harrick Diamond ConcentratIR2™ (see Figure 1), which supplies seven reflections from the sample, was installed and aligned in a commercial FTIR spectrometer. Spectra were collected at 4 cm-1 resolution and signal averaged over 64 scans. Three nail polish removers — the original, nourishing, and strengthening formulations from the same brand — were analyzed. A single drop of each type was placed on the diamond ATR crystal. Infrared spectra were collected over a range of 4000 cm-1 to 650 cm-1 periodically until the solvent, acetone, had fully evaporated.

Figure 1: ConcentratIR2™ Multiple Reflection ATR.

Results and Discussion

Many of the new peaks that emerge are present in all three samples. Only a few bands, in the 1600 cm-1 to 900 cm-1 region, as shown in Figure 2, are unique to any given sample. This is in good accordance with listed ingredients which indicates only a few unique substances in the analytes. Unique peaks include a H–C–H alkane bend at 1500 cm-1 , probably from tocopheryl acetate, an ingredient found only in the "nourishing" formula. Characteristic peaks are also observed at 1225 cm-1 and 985 cm-1 which are likely due to panthenol, another element particular to the nourishing formulation. Some other ingredients specific to one formula, such as citric acid (nourishing) and benzophenone-1 (strengthening) are present in quantities too small to be detected. Use of additional reflections or greater concentration of the sample might make these apparent.

Figure 2: ATR spectra of original (black), strengthening (red), and nourishing (blue) nail polish removers.

Conclusion

The ConcentratIR2™ can be used effectively to measure trace differences in residues of complex solutions. Since the crystal is also only 4-mm in diameter, only single drops of the sample are required, making it extremely useful in situations where only limited quantities of sample area available.

Harrick Scientific Products, Inc.

P.O. Box 277, Pleasantville, NY 10570

tel. (800) 248-3847, fax (914) 747-7209

Website: www.harricksci.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.