Mapping Molecular Mobility Pixel by Pixel

Garth Simpson of Purdue University explains how patterned photobleaching and spatial Fourier transforms can turn a single-point measurement into a diffusion coefficient at every pixel in the field of view.

Garth Simpson, who is a Tarpo Professor of Chemistry at Purdue University, is pursuing an ambitious goal: mapping molecular mobility across both space and time in complex, heterogeneous systems such as pharmaceutical formulations and materials. Here, he explains how his group is extending a well-known technique to reach it.

The starting point is fluorescence recovery after photobleaching (FRAP), which Simpson describes as a well-established, robust method for tracking molecular mobility at one location. A laser bleaches fluorophores in a spot, creating a population gradient. As bleached fluorophores diffuse out and unbleached ones diffuse in, fluorescence recovers, and the rate of recovery reveals how mobile the molecules are. The limitation is that it is a single-point measurement.

Simpson’s group instead wants to map diffusion across the entire field of view, capturing about a quarter of a million diffusion coefficients per experiment. The first step is patterning the photobleach intensity. Rather than one bleached spot, the team creates a dot array, such as a 50 × 50 grid. That yields 2,500 single-point measurements, an improvement, but still short of the goal.

To get there, the team takes a spatial Fourier transform of the image, rather than a time-dependent one, producing an image of spatial frequencies. Because a dot array in real space transforms into a dot array in Fourier space, each dot corresponds to a distinct spatial frequency. The researchers box one of those peaks, recenter it, and apply an inverse Fourier transform. The result is a real-space image whose intensity scales with the amplitude of that modulated peak.

Simpson compares the approach to lock-in detection, with the difference that the signal is mapped back into real space. By tracking how that amplitude decays over time, the team can calculate a diffusion coefficient at every pixel in the field of view. The method turns a single-point technique into a spatially resolved map of molecular mobility, offering a new window into how molecules move within complex pharmaceutical and materials systems.

Further Reading
  1. Purdue University, Garth J. Simpson. Purdue University, 2026. https://www.chem.purdue.edu/people/profile/gsimpson.html (accessed November 3, 2026).
  2. Wetzel, W.; Spectroscopy Staff. EAS 2026 to Mark 65th Anniversary with Program Focused on AI and Data Integrity. Spectroscopy Online, 2026. https://www.spectroscopyonline.com/view/eas-2026-to-mark-65th-anniversary-with-program-focused-on-ai-and-data-integrity (accessed November 3, 2026).