Application Notebook
This study describes a new automated technique for measuring protein conformation by CD spectroscopy and how changes can be followed in different pH environments. The study confirms the robustness of the technique and superb sample-to-sample reproducibility.
David Gregson and Lindsay Cole, Applied Photophysics Ltd
This study describes a new automated technique for measuring protein conformation by CD spectroscopy and how changes can be followed in different pH environments. The study confirms the robustness of the technique and superb sample-to-sample reproducibility. The improvements in productivity enable experiments to be undertaken that were previously thought to be too time consuming or laborious.
Correct protein function is critically dependent upon conformation — if a protein is not correctly folded, it will not execute its function or worse, it may be toxic. Circular dichroism (CD) spectroscopy is an exquisitely sensitive probe of protein conformation but it is often perceived as a technique limited to more specialised applications. This is to some extent because, up until now, CD spectrometers have remained resolutely manual in operation which has now been overcome by new automation technology.
Solutions of BSA, sodium phosphate, and sodium citrate were used to prepare a series of buffered protein samples from pH 2.2 to pH 8. The samples were prepared directly in a 96-well plate and each protein solution was paired with its buffer. Buffers and samples were transferred to the spectrometer using a fixed-probe x, y, z robot and CD and absorption spectra measured simultaneously and in duplicate. Each complete measurement, including washing, took 7.5 min to perform.
The baseline-corrected CD data showed that there were two phase transitions, one occurring in the range pH 2.2 to pH 3.8 and the second between pH 4 and pH 8 (Figure 1), with isosbestic points at 201 nm and 206 nm, respectively.
Figure 1: Buffer-subtracted CD spectra for BSA (260 nm ⥠λ ⥠188 nm) from pH 8 to pH 4 (left) and pH 3.8 to pH 2.2 (right). Arrows indicate increasing pH.
Plotting the raw data as a contour map gives an immediate semi-quantitative feel for the stability of the protein. Using quantitative methods, such as the global fitting of a model to the data, make it clear even to the untrained eye where regions of conformational stability exist (Figure 2). A three state model was used because the data suggest two transitions and this agrees with the well known behavior of serum albumin proteins (1). End point spectra for high-, intermediate-and low-pH were calculated as was the pKa of the transitions between them.
Figure 2: Calculated conformational stability map (CD contours). The conformation is stable at pH > 5.8.
The automated sample preparation and measurement of CD and absorption spectra have been shown to be reliable, giving quantitative results with minimum effort using robust global analysis techniques. Excellent reproducibility and sample economy together with the ability to carry out an unattended measurement at a rate of 200 samples per day considerably improves the productivity of CD measurements. These improvements offer the opportunity to study relative protein stabilities with ease, for example, in protein biotherapeutic development and clone selection procedures.
(1) D.C. Carter and J.X. Ho, "Structure of Serum Albumin." Advances in Protein Chemistry 45, 153–203. Lipoproteins, Apolipoproteins and Lipases, Academic Press 1994, Ed. Verne Schumaker. ISBN: 0-12-034245-6.
Applied Photophysics Ltd
1 Mole Business Park, Leatherhead, Surrey, United Kingdom
tel. +44 1372 386537, fax +44 1372 386477
Email: sales@photophysics.com, Website: www.photophysics.com
Getting accurate IR spectra on monolayer of molecules
April 18th 2024Creating uniform and repeatable monolayers is incredibly important for both scientific pursuits as well as the manufacturing of products in semiconductor, biotechnology, and. other industries. However, measuring monolayers and functionalized surfaces directly is. difficult, and many rely on a variety of characterization techniques that when used together can provide some degree of confidence. By combining non-contact atomic force microscopy (AFM) and IR spectroscopy, IR PiFM provides sensitive and accurate analysis of sub-monolayer of molecules without the concern of tip-sample cross contamination. Dr. Sung Park, Molecular Vista, joined Spectroscopy to provide insights on how IR PiFM can acquire IR signature of monolayer films due to its unique implementation.
Researchers Develop Adaptive Gap-Tunable SERS Device
April 24th 2024In a new study, researchers from Ulsan National Institute of Science and Technology and Pohang University of Science and Technology presented a new surface-enhanced Raman spectroscopy (SERS) device, improving gap plasmon resonance.