More Raman Analyses Less Fluorescence Interference

In recent years, the spectroscopy community has observed rapid development of Raman instrumentation and its usefulness in a variety of applications. Routine Raman analysis with 785 nm excitation has served well for the great majority of industrial applications and has become the most favored instrument configuration.

Duyen Nguyen and Eric Wu, Enwave Optronics, Inc.

In recent years, the spectroscopy community has observed rapid development of Raman instrumentation and its usefulness in a variety of applications. Routine Raman analysis with 785 nm excitation has served well for the great majority of industrial applications and has become the most favored instrument configuration. Nevertheless, the need still persists for further reduction of fluorescence interference in many important applications containing fluorescent samples.

Continuing the success of the ProRaman Series, Enwave Optronics recently announced the implementation of the ProRaman-L9 Series instruments with excitation wavelength around 900 nm for analysis of materials prone to fluorescence interference. The deployment of ProRaman-L9 has proven to be invaluable for various applications requiring further fluorescence reduction from using 785 nm excitation.

Experiment

The samples chosen for this experiment were red wine and wheat flour, both of which were previously reported unable to be analyzed by 785 nm Raman instruments (due to fluorescence). The samples were measured using ProRaman-L-785 and ProRaman-L9. Both of the ProRaman instruments were equipped with a fiber optics probe for sample interface. The red wine sample was kept in a quartz cuvette during data collection. Whole wheat flour was measured using a sapphire lens probe for direct measurement with the sample in a plastic bag.

Results and Discussion

Figure 1 shows the comparison between Raman spectra of red wine measured with Proraman-L-785 and ProRaman-L9. With 785 nm excitation, fluorescence quickly filled the intensity dynamic range, leaving little room for Raman signal to be analyzed. With ProRaman-L9, the Raman to fluorescence ratio improved about 10 times and after simple baseline subtraction, the ethanol Raman peaks appeared strong and the analysis could be readily performed.

Figure 1: Raman spectra of red wine measured with ProRaman-L-785 and ProRaman-L9. Baseline subtracted ProRaman-L9 data shows in brown trace.

Figure 2 shows the measurements of whole wheat flour sample measured with ProRaman-L-785 and ProRaman-L9. Again, obvious improvement of the Raman to fluorescence ratio can be seen, and the baseline-subtracted spectrum of the whole wheat flour could be readily analyzed.

Figure 2: Raman spectra of whole wheat flour measured with ProRaman-L-785 and ProRaman –L9. Baseline subtracted ProRaman-L9 data shows in brown trace.

Conclusion

Utilizing longer wavelengths into the near infrared for Raman excitation further reduces the interference of fluorescence and improves the capability for Raman analysis in samples previously thought tough to measure. Red wine and whole wheat flour are two excellent examples showing the improved usefulness and capability for Raman analysis using ProRaman-L9. Other applications such as motor oil, pharmaceuticals, etc., may also benefit from this longer excitation wavelength Raman system.

Enwave Optronics, Inc.

18200 McDurmott St., Suite A, Irvine, CA 92614

tel. (949) 955-0258, fax (949) 955-0259

Website: www.enwaveopt.com


Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Healthcare administrator reviewing an insurance policy schedule with a stethoscope on a clipboard, health insurance coverage, medical underwriting, prior authorization, and patient claims processing. | Image Credit: Joi_studio - stock.adobe.com
In a SciX interview, Marc Porter covered a lot of ground, from a student-driven pivot into immunoassays to a baseball analogy for Stokes and anti-Stokes scattering, and shares the open questions he is still chasing.
Highlights from What’s Nu October 2026
In this Q&A, we recap the October edition of “What’s Nu,” which covered topics such as how novel spectroscopic and analytical methods are transforming environmental monitoring, medical treatment, and accessibility in scientific research.