Supplementary Information: Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes
This information is a supplementary to the article “Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes” that was published in the June 2016 Spectroscopy supplemental issue Raman Technology for Today’s Spectroscopists (1).
This information is supplementary to the article “Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes” that was published in the June 2016 Spectroscopy supplemental issue Raman Technology for Today’s Spectroscopists (1).
Optical Anisotropy Calculation in Figure 4
One way of using the Raman imaging data is to obtain an average reading for a larger area of the sample compared to a single point measurement. The boxed areas in Figure 4 represent areas where the Raman spectral measurements were averaged to produce a single Raman spectrum. The boxed area in Figure 4 on the left is an area of unaligned SWCNTs; the boxed area on the right is an area of aligned SWCNTs.
Each boxed area in Figure 4 is 11 x 47 µm (517 sq. µm) with spectra collected at a spatial resolution (pixel size) of 0.3 µm; this comes out to ~1800 individual spectra. Therefore, each boxed area produces a single Raman spectrum representing the averaged sample area. We calculated the peak area of the G-Band from the averaged spectrum collected at each polarization configuration (V_V, H_H) and took the ratio. A correction factor was applied to compensate for the reduced optical throughput in the H_H configuration. This correction factor is calculated from the silicon peak at 519 cm-1 Raman shift which is not sensitive to polarization.
Aligned s-SWCNTs Region:
Randomly-Oriented s-SWCNTs Region:
Therefore, the average optical anisotropy value for the aligned s-SWCNTs is ≈22 and for randomly oriented s-SWCNTs is ≈1.4.
Orientation Image in Figure 5
The V_V and H_H polarization images were exported from OMNICxi as grayscale TIFF images having the same intensity scale. This preserved the relative numerical values of each pixel between the two images. The TIFF images were imported into ImageJ where the subsequent orientation image was created by dividing the V_V image by the H_H image (3). The resulting image was scaled by a factor of 0.882 to correct for the lower instrument throughput caused by additional polarization optics (half wave plate) when collecting the H_H image.
References
- A. Mashal, D. Wieboldt, K. Jinkins, and M.S. Arnold, Spectroscopy31(s6), 36–44 (2016).
- Y. Joo, G. Brady, M. Arnold, and P. Gopalan, American Chemical Society, 3460–3466 (2014).
- J. Schindelin, et al., Nature Methods9(7), 676–682, PMID 22743772 (2012).
Articles in this issue
over 10 years ago
Evaluation of Eye Shadow Compounds Using Raman Microspectroscopyover 10 years ago
Using SERS to Study How Cells Respond to Pharmaceuticalsover 10 years ago
Tech Tip: Choosing Raman Excitation Wavelengthsover 10 years ago
Vol 31 No s6 Spectroscopy June 2016 Raman Technology Supplement PDFRelated to this article