Supplementary Information: Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes

This information is a supplementary to the article “Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes” that was published in the June 2016 Spectroscopy supplemental issue Raman Technology for Today’s Spectroscopists (1).

This information is supplementary to the article “Polarized Raman Spectroscopy of Aligned Semiconducting Single-Walled Carbon Nanotubes” that was published in the June 2016 Spectroscopy supplemental issue Raman Technology for Today’s Spectroscopists (1). 

 

Optical Anisotropy Calculation in Figure 4

 

One way of using the Raman imaging data is to obtain an average reading for a larger area of the sample compared to a single point measurement. The boxed areas in Figure 4 represent areas where the Raman spectral measurements were averaged to produce a single Raman spectrum. The boxed area in Figure 4 on the left is an area of unaligned SWCNTs; the boxed area on the right is an area of aligned SWCNTs.

Each boxed area in Figure 4 is 11 x 47 µm (517 sq. µm) with spectra collected at a spatial resolution (pixel size) of 0.3 µm; this comes out to ~1800 individual spectra. Therefore, each boxed area produces a single Raman spectrum representing the averaged sample area. We calculated the peak area of the G-Band from the averaged spectrum collected at each polarization configuration (V_V, H_H) and took the ratio. A correction factor was applied to compensate for the reduced optical throughput in the H_H configuration. This correction factor is calculated from the silicon peak at 519 cm-1 Raman shift which is not sensitive to polarization.

Aligned s-SWCNTs Region:

 

 

 

 

 

 

Randomly-Oriented s-SWCNTs Region:

Therefore, the average optical anisotropy value for the aligned s-SWCNTs is ≈22 and for randomly oriented s-SWCNTs is ≈1.4.

 

Orientation Image in Figure 5

 

The V_V and H_H polarization images were exported from OMNICxi as grayscale TIFF images having the same intensity scale. This preserved the relative numerical values of each pixel between the two images. The TIFF images were imported into ImageJ where the subsequent orientation image was created by dividing the V_V image by the H_H image (3). The resulting image was scaled by a factor of 0.882 to correct for the lower instrument throughput caused by additional polarization optics (half wave plate) when collecting the H_H image.

 

References

  • A. Mashal, D. Wieboldt, K. Jinkins, and M.S. Arnold, Spectroscopy31(s6), 36–44 (2016).

  • Y. Joo, G. Brady, M. Arnold, and P. Gopalan, American Chemical Society, 3460–3466 (2014).

  • J. Schindelin, et al., Nature Methods9(7), 676–682, PMID 22743772 (2012).

Related to this article

Monitoring Sugar Consumption and Ethanol Production During Yeast Fermentation Using a Raman Plate Reader and Multivariate Analysis
This column shows that a Raman plate reader paired with partial least squares regression can simultaneously quantify glucose, fructose, sucrose, and ethanol in yeast fermentation samples, using turbidity-matched calibration standards to correct for yeast-induced turbidity, and offers a rapid, label-free, high-throughput method for tracking sugar consumption and ethanol production in strain screening, media optimization, and process development.
Highly Substituted Benzene Rings—The Rest of the Story II: Spectra
This column, the second installment on highly substituted benzene rings, shows that tri- and tetra-substituted substitution patterns can be distinguished by whether the ring bending peak is present and by the position of the in-phase aryl C-H wagging peak, which shifts according to how many adjacent hydrogens the ring carries.